  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92061",
      "recalling_firm": "CareFusion 303, Inc.",
      "address_1": "10020 Pacific Mesa Blvd",
      "address_2": "N/A",
      "postal_code": "92121-4386",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of KS, NY, NC, LA, MA, MI.",
      "recall_number": "Z-1579-2023",
      "product_description": "BD Pyxis MedStation ES (Med ES Main), REF: 323; BD Pyxis MedStation ES (Med ES Main Tower), REF: 352; BD Pyxis Anesthesia Station ES, REF: 327; BD Pyxis CII Safe ES Tower Main, REF: 1116-00; BD Pyxis CII Safe ES, Desktop PC, W-LESS SCNR, BIO, REF: 107-255-01",
      "product_quantity": "1,474",
      "reason_for_recall": "Automated dispensing cabinet software is experiencing: 1) ES device download failure resulting in partial patient information loss that leads to 2) partial loss of patient transaction data, seen as a server upload processing error. Outdated/incorrect information display may contribute to the removal of medication to which the patient is allergic, incorrect dose amount, or discontinued medication.",
      "recall_initiation_date": "20230405",
      "center_classification_date": "20230511",
      "report_date": "20230517",
      "code_info": "REF/UDI-DI/Software: 323/10885403512667/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4; 352/10885403512674/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4; 327/10885403477836/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4; 1116-00/10885403512605/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4; 107-255-01/1.7.0, 1.7.1, 1.7.2, 1.7.3, 1.7.4"
    }
  ]
}