  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Boulder",
      "state": "CO",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73701",
      "recalling_firm": "Medtronic",
      "address_1": "6135 Gunbarrel Ave",
      "address_2": "N/A",
      "postal_code": "80301-3214",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. Product was also distributed worldwide. There is 1 VA customer.",
      "recall_number": "Z-1577-2016",
      "product_description": "Battery Pack Used with the Capnostream\"20 and Capnostream\"20p Patient Monitor, Battery Packs Model Number - 016400, 010520.  Battery Pack used with the Capnostream\"20 and Capnostream\"20p Patient Monitor. The battery pack powers the monitor when AC power is not available or when monitoring is required during intra-hospital transfer of patients. The Capnostream\"20 and Capnostream\"20p is a combined capnograph/pulse oximeter monitor and its accessories are intended to provide professionally trained health care providers with continuous, non-invasive measurement and monitoring of carbon dioxide concentration of the expired and inspired breath and respiration rate, and with continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. It is intended for use with neonatal, pediatric, and adult patients in hospitals, hospital-type facilities, and intra-hospital transport environments",
      "product_quantity": "9,817 batteries (8,170 with monitors and 1,647 as spare parts)",
      "reason_for_recall": "Medtronic Respiratory & Monitoring Solutions Announces a Voluntary Recall of a Battery Pack Used with the Capnostream\"20 and Capnostream\"20p Patient Monitor Due to the Battery Pack Might Experience Thermal Runaway with the Potential of Fire Risk to the Monitor.",
      "recall_initiation_date": "20160413",
      "center_classification_date": "20160512",
      "termination_date": "20171017",
      "report_date": "20160518",
      "code_info": "Product # 016400; All SNs, irrespective of production date. Product # 010520; All SNs with production date starting 14.14"
    }
  ]
}