  "results": [
    {
      "status": "Terminated",
      "city": "Vernon Hills",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78412",
      "recalling_firm": "Vyaire Medical",
      "address_1": "75 N Fairway Dr",
      "address_2": "N/A",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) and in the countries of  Canada, Puerto Rico, United Arab Emirates, Philippines, and  Europe.",
      "recall_number": "Z-1575-2018",
      "product_description": "AirLife Resuscitation, Pediatric, with mask, oxygen reservoir bag, manometer, pressure-relief valve, PEEP valve    Product Usage:  The AirLife Resuscitation devices are manual resuscitation devices that are used to provide positive pressure ventilation to patients who are not breathing or cannot adequately breathe on their own.",
      "product_quantity": "270 units",
      "reason_for_recall": "The mask component on various lots of the resuscitation devices have been identified as having the  potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the  mask from the elbow of the resuscitator.",
      "recall_initiation_date": "20170427",
      "center_classification_date": "20180510",
      "termination_date": "20190509",
      "report_date": "20180516",
      "code_info": "Part Number: 2K8039M. Lots: 0001000959, 0001002931, 0001007291, 0001016628, 0001019907"
    }
  ]
}