      "status": "Ongoing",
      "city": "Memphis",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92142",
      "recalling_firm": "Medtronic Sofamor Danek USA Inc",
      "address_1": "1800 Pyramid Pl",
      "address_2": "N/A",
      "postal_code": "38132-1703",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US:  IN, CT, NE",
      "recall_number": "Z-1574-2023",
      "product_description": "Distractor Rack, REF 5598007, which may be packaged in Modular Distractor Set, Loaner Kit Number SPS03139; for use in spinal surgery",
      "product_quantity": "4 units",
      "reason_for_recall": "Two lots of the 5598007 Distractor Rack which were not manufactured to specification. The crest width of the gear is too wide and may cause difficulty or the inability to operate the Distractor.",
      "recall_initiation_date": "20211119",
      "center_classification_date": "20230510",
      "report_date": "20230517",
      "code_info": "UDI/DI (GTIN) 00763000401535, Lot Numbers EM21D006, EM21D007;  Loaner Kit Lot numbers:  0072, 0073, 0077, 0098, 0103"
    }
  ]
}