      "status": "Terminated",
      "city": "Tempe",
      "state": "AZ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84914",
      "recalling_firm": "Bard Peripheral Vascular Inc",
      "address_1": "1625 W 3rd St Bldg 1",
      "address_2": "N/A",
      "postal_code": "85281-2438",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS,KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT,VA, VT, WA, WI, WV, WY and countries of Brazil, Canada, EMEA, Korea, Mexico, New Zealand.",
      "recall_number": "Z-1572-2020",
      "product_description": "1. Power Port isp M.R.I. Implantable Port with Attachable 9.6 F Open-Ended Single-Lumen Venous Catheter with Suture Plugs, Ref: 1809660, UDI: (01)00801741027154  2. PowerPort isp M.R.I. Implantable Port With Attachable 9.6 F Open-Ended Single-Lumen Venous Catheter Without Suture Plugs, Ref: 1809661, UDI: (01)00801741027161  3. PowerPort isp M.R.I. Implantable Port with Pre-Attached 9.6F Open-Ended Single-Lumen Venous Catheter with Suture Plugs, Ref: 1859660, UDI: (01)00801741027222  4. PowerPort isp M.R.I. Implantable Port with Pre-Attached 9.6 F Open-Ended Single-Lumen Venous Catheter without Suture Plugs, Ref: 1859661, UDI: (01)00801741027239    Product Usage: Bard implantable ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.",
      "product_quantity": "824 Catheters",
      "reason_for_recall": "Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a  9.6Fr catheter.  This can result in a prolongation of operation.",
      "recall_initiation_date": "20191001",
      "center_classification_date": "20200326",
      "termination_date": "20220218",
      "report_date": "20200401",
      "code_info": "1. Lot # for 1809660 - RECT0092  2. Lot # for 1809661 - RECT0190  3. Lot # for 1859660 - RECU0481, RECR0983, RECR1885, RECR2220, RECS0030, RECS2758  4. Lot # for 1859661 - RECS2974",
      "more_code_info": ""
    }
  ]
}