  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mountain View",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73886",
      "recalling_firm": "Siemens Medical Solutions USA, Inc.",
      "address_1": "685 E Middlefield Rd",
      "address_2": "N/A",
      "postal_code": "94043-4045",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Distributed in Germany, U.A.E., Hungary and Brazil.",
      "recall_number": "Z-1572-2016",
      "product_description": "ACUSON X700 Ultrasound System with software version 1.0.04.  Model numbers:  10658844 \u0013 ACUSON X700 Ultrasound System  10658846 - ACUSON X700 Ultrasound System (Russia)  10658845 - ACUSON X700 Ultrasound System (Korea)    Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the following applications: Cardiac (Adult, Pediatric), Transesphageal (Cardiac), Intracardiac, Cerebrovascular, Peripheral Vessel, Abdominal, Renal, Fetal, Abdominal, Intra-operative, Pediatric, Small Organ, Neonatal Cepahalic, Adult Cephalic, Orthopedics, Musculo-skeletal Conventional, Musculo-skeletal Superficial, Pelvic, Obstetrical, Gynecological and Urological applications.  The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.",
      "product_quantity": "4",
      "reason_for_recall": "Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which requires a reboot to recover system operation.",
      "recall_initiation_date": "20150905",
      "center_classification_date": "20160427",
      "termination_date": "20160427",
      "report_date": "20160504",
      "code_info": "Material (System)      IVK Group                   Serial number:  10658844               ACUSON X700           356559  10658844               ACUSON X700           356733  10658844               ACUSON X700           356734  10658844               ACUSON X700           358714"
    }
  ]
}