{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lake Forest",
      "state": "IL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "91874",
      "recalling_firm": "ICU Medical Inc",
      "address_1": "600 N Field Dr",
      "address_2": "N/A",
      "postal_code": "60045-4835",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "worldwide",
      "recall_number": "Z-1567-2023",
      "product_description": "ASM Replacement Battery, Component Number SUB0000864",
      "product_quantity": "9984 units",
      "reason_for_recall": "Due to a manufacturing defect from the battery supplier, affected batteries may experience a loss of capacity earlier in the battery lifecycle than expected and overall battery runtime may decrease earlier in the battery lifecycle than expected.",
      "recall_initiation_date": "20230322",
      "center_classification_date": "20230516",
      "report_date": "20230524",
      "code_info": "Lot Numbers:  6036488, 6827738, 6827739, 6827740, 8408750, 8408755, 8408757"
    }
  ]
}