{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Burlingame",
      "address_1": "863 Mitten Rd Ste 100A",
      "reason_for_recall": "Slow deflation time: Slower than normal balloon deflation time of greater than 10 seconds.  The balloon normally deflates in less than 5 seconds.  The device performance specification for deflation time is 10 seconds maximum.",
      "address_2": "",
      "product_quantity": "2379 units",
      "code_info": "All lot numbers lower than PLN00749 are subject to this recall.",
      "center_classification_date": "20130621",
      "distribution_pattern": "Worldwide Distribution-US Distribution in New Jersey and the countries of Germany, Switzerland, Italy, Norway, The Netherlands, Spain, Portugal, Ireland, and Austria.",
      "state": "CA",
      "product_description": "TRUE Dilatation Balloon Valvuloplasty Catheter,     REF Numbers: 02045-11, 02245-12, 02445-12, and 02645-13; Sizes:   20mm x 4.5cm, 22mm x 4.5cm, 24mm x 4.5cm, and 26mm x 4.5cm;     Intended to be used for balloon aortic valvuloplasty.",
      "report_date": "20130703",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Loma Vista Medical",
      "recall_number": "Z-1566-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65343",
      "termination_date": "20130717",
      "more_code_info": "",
      "recall_initiation_date": "20130318",
      "postal_code": "94010-1307",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}