  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Beverly",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63145",
      "recalling_firm": "SpineFrontier, Inc.",
      "address_1": "500 Cummings Ctr Ste 3500",
      "address_2": "N/A",
      "postal_code": "01915-6516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide - OR, FL, MA, TX, CO, OK, IL, and  MO",
      "recall_number": "Z-1565-2014",
      "product_description": "SpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in spinal fusion surgery.  Product Usage: The S-Lift Lock Shaft SI50001 mates with SI50002, Implant Inserter, to aid in securing the implant to the whole implant inserter assembly, SI50000.",
      "product_quantity": "30 devices",
      "reason_for_recall": "It has been verified through field use that there are two possible breakages that may occur when using SI50001 Revision D Lock Shaft Inserter, possibly resulting in instrument fragments entering the sterile field:  Impact Cap: The impact cap on the Lock Shaft Inserter can shear off under high-impact force during insertion and removal.  Pin: The pin used to guide the Lock Shaft Inserter into the ma",
      "recall_initiation_date": "20111223",
      "center_classification_date": "20140507",
      "termination_date": "20140520",
      "report_date": "20140514",
      "code_info": "S-Lift Lock Shaft Part number; SI50001  Revisions: A, B  Lots: 42613, 2478901R, 42615, 7625701"
    }
  ]
}