  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "East Walpole",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67839",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc",
      "address_1": "333 Coney St",
      "address_2": "N/A",
      "postal_code": "02032-1516",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of Austria Belgium Bulgaria Cyprus Czech Republic Denmark Estonia Finland Finland France Fren.Polynesia Germany Great Britain Greece Guadeloupe Hungary Iceland Ireland Israel Italy Latvia Lithuania Luxembourg Malta Martinique Netherlands Norway Poland Portugal Qatar Reunion Romania Russian Fed. Serbia Slovakia Slovenia Spain Sweden Switzerland Turkey Unit.Arab Emir.",
      "recall_number": "Z-1564-2014",
      "product_description": "Siemens ADVIA Centaur Folate   Cat No. 06891541 SMN 10325366 (500 tests)  Cat No. 06367974 SMN 10310308 (100 tests)  Cat No. 00203473 SMN 10331250 (Ref) (500 tests)  Cat No. 09132781 SMN 10340209 (Ref) (2500 tests)    For in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the ADVIA Centaur and ADVIA Centaur XP systems",
      "product_quantity": "49123 units",
      "reason_for_recall": "Serum controls may go out of range low with the ADVIA Centaur ¿Systems Folate assay and a shift in whole blood patient may result with kit lots ending in 219 and 223",
      "recall_initiation_date": "20140325",
      "center_classification_date": "20140507",
      "termination_date": "20150430",
      "report_date": "20140514",
      "code_info": "Kit Lots ending in and exp date:  219   Exp. Date 25 March, 2014;  222   Exp Date 12 May, 2014;  224   Exp Date 04 August, 2014;  225  Exp Date 04 August, 2014;  226  Exp Date 12 September, 2014;  227  Exp. Date  06 October, 2014"
    }
  ]
}