  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87765",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: IN, MA, CA, NJ, TX, NE, NY, VA, MD, MN, OR, IL, CT, DC, MS. Foreign: Albania, Argentina, Australia, Bangladesh, Bolivia, Bosnia & Herzegovina, Cameroon, Canada, Colombia, Cyprus, Denmark, Egypt, Estonia, Fiji, France, Germany, Greece, Hong Kong, Iraq, Italy, Ireland, Jordan, Kazakhstan, Kenya, Lithuania, Luxembourg, Mauritius, Mexico, Myanmar, Nepal, New Zealand, Pakistan, Palestine, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Thailand, The Netherlands, Turkey, UK, Ukraine, USA, and Vietnam.",
      "recall_number": "Z-1563-2021",
      "product_description": "Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator.   Catalogue Number: CAL2351",
      "product_quantity": "98 kits",
      "reason_for_recall": "Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.",
      "recall_initiation_date": "20210329",
      "center_classification_date": "20210512",
      "termination_date": "20211122",
      "report_date": "20210519",
      "code_info": "Batch / Lot number: 1162UE  GTIN: 05055273200966",
      "more_code_info": ""
    }
  ]
}