  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Saint Paul",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73117",
      "recalling_firm": "3M Company - Health Care Business",
      "address_1": "3m Center 2510 Conway Ave , B# 275-5-W-6",
      "address_2": "N/A",
      "postal_code": "55144-0001",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "AL, CO, CT, FL, GA, IN, NC, NJ, NY, OK, OR, PA, RI, VA, WI, and OUS to include:  Argentina, Canada, Chile, China, Columbia, El Salvador, Germany, India, Indonesia, Japan, Korea, Mexico, Peru, Philippines, South Africa, Switzerland, Thailand, Trinidad, Turkey, United Arab Emirates, Vietnam.",
      "recall_number": "Z-1563-2016",
      "product_description": "3M Steri-Vac Gas Sterilizer/Aerator,  models 5XL and 8XL.",
      "product_quantity": "74 valves",
      "reason_for_recall": "3M discovered a valve with incorrectly labeled \"input\" and \"output\" orientation.  This defect can result in the valve being installed backwards, creating a remote possibility of valve failure.  A failure may result in EO being vented outside the sealed chamber during a limited period in the EO injection phase of the cycle.",
      "recall_initiation_date": "20160122",
      "center_classification_date": "20160424",
      "termination_date": "20160825",
      "report_date": "20160504",
      "code_info": "Model 5XL, Serial Numbers:  722612 - 722683, 820352 - 820360    Model 8XL, Serial Numbers:  351303 - 351399, 450144 - 450147"
    }
  ]
}