      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73617",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- US, to the states of AZ, CA, DC, FL, GA, ID, IL IN, KY, MD, MI, MN, MO, NY, NC, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI; and, to the countries of AUSTRALIA, BELGIUM. BOLIVIA, CANADA, CHILE, DENMARK, FINLAND, FRANCE, GERMANY, HUNGARIA, INDIA, IRELAND, JAPAN, KAZAKHSTAN, KOREA, KUWAIT, NETHERLANDS, NEW ZEALAND, SAUDI ARABIA, SPAIN, SWITZERLAND, THAILAND, TURKEY, UNITED ARAB EMIRATES, and UNITED KINGDOM.",
      "recall_number": "Z-1562-2016",
      "product_description": "GE Healthcare, Aisys and Aisys CS2.    Intended to provide general inhalation anesthesia and ventilator support to a wide range of patients (neonatal, pediatric, and adult).",
      "product_quantity": "952 (488 US; 464 OUS)",
      "reason_for_recall": "The vacuum suction tubing could be kinked within the machine resulting in suction less than the 20 lpm required by ISO 10079-3. This could result in inadequate suction performance to clear the airway and inadequate ventilation of the patient.",
      "recall_initiation_date": "20160315",
      "center_classification_date": "20160422",
      "termination_date": "20161107",
      "report_date": "20160504",
      "code_info": "Mfg Lot or Serial # UDI IDENTIFIER  APWU00723 not applicable  APWU00724 not applicable  APWU01988 010084068210229221APWU0198811150921  APWU01989 010084068210229221APWU0198911150921  APWU01990 010084068210229221APWU0199011150921  APWU01991 010084068210229221APWU0199111150921  APWU01992 010084068210229221APWU0199211150921  APWU01993 010084068210229221APWU0199311150921  APWU01994 010084068210229221APWU0199411150921  APWU01995 010084068210229221APWU0199511150921  APWU01996 010084068210229221APWU0199611150921  APWU01997 010084068210229221APWU0199711150921  APWU00707 not applicable  APWU00708 not applicable  APWU00709 not applicable  APWU00710 not applicable  APWU00711 not applicable  APWU02166 010084068210229221APWU0216611151014  APWU02170 010084068210229221APWU0217011151014  APWU02171 010084068210229221APWU0217111151014  APWU02172 010084068210229221APWU0217211151014  APWU00712 not applicable  APWU00505 not applicable  APWU00506 not applicable  APWU00507 not applicable  APWU00474 not applicable  APWU00475 not applicable  APWU00476 not applicable  APWU00477 not applicable  APWU00478 not applicable  APWU00479 not applicable  APWU01486 not applicable  APWU01487 not applicable  APWU01488 not applicable  APWU01489 not applicable  APWU01490 not applicable  APWU01491 not applicable  APWU01492 not applicable  APWU01493 not applicable  APWU01494 not applicable  APWU01495 not applicable  APWU00611 not applicable  APWU00612 not applicable  APWU00613 not applicable  APWU00614 not applicable  APWU02163 010084068210229221APWU0216311151014  APWU02164 010084068210229221APWU0216411151014  APWU02165 010084068210229221APWU0216511151014  APWU02051 010084068210229221APWU0205111151002  APWU01909 010084068210229221APWU0190911150902  APWU02017 010084068210229221APWU0201711150921  APWU02018 010084068210229221APWU0201811150921  APWU00297 not applicable  APWU01608 not applicable  APWU01609 not applicable  APWU01610 not applicable  APWU01611 not applicable  APWU01612 not applicable  APWU01613 not applicable  APWU01614 not applicable  APWU01615 not applicable  APWU00579 not applicable  APWU00580 not applicable  APWU00581 not applicable  APWU00582 not applicable  APWU00583 not applicable  APWU00584 not applicable  APWU00585 not applicable  APWU00586 not applicable  APWU00587 not applicable  APWU00588 not applicable  APWU00589 not applicable  APWU00590 not applicable  APWU00591 not applicable  APWU00592 not applicable  APWU00593 not applicable  APWU00594 not applicable  APWU00595 not applicable  APWU00596 not applicable  APWU00597 not applicable  APWU00598 not applicable  APWU00599 not applicable  APWU00600 not applicable  APWU00601 not applicable  APWU00602 not applicable  APWU00603 not applicable  APWU00728 not applicable  APWU00729 not applicable  APWU00730 not applicable  APWU00751 not applicable  APWU00752 not applicable  APWU00753 not applicable  APWU00754 not applicable  APWU00755 not applicable  APWU00604 not applicable  APWU00605 not applicable  APWU00606 not applicable  APWU00713 not applicable  APWU01410 not applicable  APWU01411 not applicable  APWU01412 not applicable  APWU02108 010084068210229221APWU0210811151007  APWU00293 not applicable  APWU00260 not applicable  APWU01616 not applicable  APWU01617 not applicable  APWU01618 not applicable  APWU01619 not applicable  APWU01620 not applicable  APWU01621 not applicable  APWU01622 not applicable  APWU01623 not applicable  APWU01699 not applicable  APWU01700 not applicable  APWU01701 not applicable  APWU01702 not applicable  APWU01703 not applicable  APWU01704 not applicable  APWU01705 not applicable  APWU01706 not applicable  APWU01707 not applicable  APWU01708 not applicable  APWU01709 not applicable  APWU01710 not applicable  APWU01652 not applicable  APWU01653 not applicable  APWU01654 not applicable  APWU01655 not applicable  APWU01096 not applicable  APWU01097 not applicable  APWU01098 not applicable  APWU01099 not applicable  APWU01100 not applicable  APWU01101 not applicable  APWU01102 not applicable  APWU01103 not applicable  APWU01104 not applicable  APWU01105 not applicable  APWU01106 not applicable  APWU01107 not applicable  APWU01108 not applicable  APWU01109 not applicable  APWU01736 not applicable  APWU01737 not applicable  APWU00864 not applicable  APWU00865 not applicable  APWU00866 not applicable  APWU00867 not applicable  APWU00868 not applicable  APWU00869 not applicable  APWU00870 not applicable  APWU00871 not applicable  APWU00607 not applicable  APWU00608 not applicable  APWU00610 not applicable  APWU02176 010084068210229221APWU0217611151014  APWU01910 010084068210229221APWU0191011150902  APWU00913 not applicable  APWU01911 010084068210229221APWU0191111150902  APWU01912 010084068210229221APWU0191211150902  APWU01913 010084068210229221APWU0191311150902  APWU01914 010084068210229221APWU0191411150902  APWU01929 010084068210229221APWU0192911150917  APWU01930 010084068210229221APWU0193011150917  APWU01931 010084068210229221APWU0193111150917  APWU01932 010084068210229221APWU0193211150917  APWU01933 010084068210229221APWU0193311150917  APWU01934 010084068210229221APWU0193411150917  APWU01935 010084068210229221APWU0193511150917  APWU01936 010084068210229221APWU0193611150917  APWU01937 010084068210229221APWU0193711150917  APWU01938 010084068210229221APWU0193811150917  APWU00764 not applicable  APWU00765 not applicable  APWU00766 not applicable  APWU00889 not applicable  APWU01443 not applicable  APWU01444 not applicable  APWU01445 not applicable  APWU01446 not applicable  APWU01447 not applicable  APWU01448 not applicable  APWU01449 not applicable  APWU01450 not applicable  APWU01451 not applicable  APWU01452 not applicable  APWU00725 not applicable  APWU00726 not applicable  APWU00727 not applicable  APWU00241 not applicable  APWU00242 not applicable  APWU00243 not applicable  APWU00244 not applicable  APWU00245 not applicable  APWU00486 not applicable  APWU00487 not applicable  APWU01665 not applicable  APWU01666 not applicable  APWU01667 not applicable  APWU01668 not applicable  APWU01669 not applicable  APWU01670 not applicable  APWU01671 not applicable  APWU01672 not applicable  APWU01673 not applicable  APWU01674 not applicable  APWU01675 not applicable  APWU01684 not applicable  APWU01685 not applicable  APWU01686 not applicable  APWU01687 not applicable  APWU01688 not applicable  APWU01689 not applicable  APWU01690 not applicable  APWU01691 not applicable  APWU01692 not applicable  APWU01693 not applicable  APWU01694 not applicable  APWU01540 not applicable  APWU01541 not applicable  APWU01542 not applicable  APWU01543 not applicable  APWU01544 not applicable  APWU01545 not applicable  APWU01546 not applicable  APWU01547 not applicable  APWU01548 not applicable  APWU01549 not applicable  APWU01560 not applicable  APWU01561 not applicable  APWU01562 not applicable  APWU01563 not applicable  APWU01564 not applicable  APWU01565 not applicable  APWU01566 not applicable  APWU01567 not applicable  APWU01568 not applicable  APWU01569 not applicable  APWU01570 not applicable  APWU00937 not applicable  APWU00938 not applicable  APWU02053 010084068210229221APWU0205311151002  APWU02054 010084068210229221APWU0205411151002  APWU02055 010084068210229221APWU0205511151002  APWU02056 010084068210229221APWU0205611151002  APWU02057 010084068210229221APWU0205711151002  APWU02058 010084068210229221APWU0205811151002  APWU00540 not applicable  APWU00541 not applicable  APWU00542 not applicable  APWU00543 not applicable  APWU00621 not applicable  APWU00622 not applicable  APWU00623 not applicable  APWU00624 not applicable  APWU00625 not applicable  APWU00626 not applicable  APWU00627 not applicable  APWU00628 not applicable  APWU00629 not applicable  APWU00630 not applicable  APWU00631 not applicable  APWU00632 not applicable  APWU00633 not applicable  APWU00634 not applicable  APWU00635 not applicable  APWU00636 not applicable  APWU00637 not applicable  APWU00638 not applicable  APWU00639 not applicable  APWU00645 not applicable  APWU00646 not applicable  APWU00647 not applicable  APWU00648 not applicable  APWU00649 not applicable  APWU00650 not applicable  APWU00651 not applicable  APWU00652 not applicable  APWU00653 not applicable  APWU00654 not applicable  APWU00655 not applicable  APWU00656 not applicable  APWU00657 not applicable  APWU00658 not applicable  APWU00659 not applicable  APWU00660 not applicable  APWU00661 not applicable  APWU00662 not applicable  APWU00663 not applicable  APWU00664 not applicable  APWU00665 not applicable  APWU00666 not applicable  APWU00667 not applicable  APWU00668 not applicable  APWU00669 not applicable  APWU00670 not applicable  APWU00671 not applicable  APWU00672 not applicable  APWU00673 not applicable  APWU00674 not applicable  APWU00675 not applicable  APWU00676 not applicable  APWU00677 not applicable  APWU00678 not applicable  APWU00250 not applicable  APWU00252 not applicable  APWU00517 not applicable  APWU00519 not applicable  APWU00520 not applicable  APWU00524 not applicable  APWU00274 not applicable  APWU00275 not applicable  APWU00276 not applicable  APWU00277 not applicable  APWU00278 not applicable  APWU00279 not applicable  APWU00280 not applicable  APWU00281 not applicable  APWU00282 not applicable  APWU00283 not applicable  APWU00284 not applicable  APWU00285 not applicable  APWU00286 not applicable  APWU00287 not applicable  APWU00288 not applicable  APWU00289 not applicable  APWU00290 not applicable  APWU00291 not applicable  APWU00292 not applicable  APWU01360 not applicable  APWU01361 not applicable  APWU01362 not applicable  APWU01363 not applicable  APWU01364 not applicable  APWU01365 not applicable  APWU01366 not applicable  APWU01367 not applicable  APWU01368 not applicable  APWU01369 not applicable  APWU01370 not applicable  APWU00485 not applicable  APWU01348 not applicable  APWU01349 not applicable  APWU01350 not applicable  APWU01351 not applicable  APWU01352 not applicable  APWU01353 not applicable  APWU01354 not applicable  APWU01355 not applicable  APWU01356 not applicable  APWU01357 not applicable  APWU01358 not applicable  APWU01331 not applicable  APWU01332 not applicable  APWU01333 not applicable  APWU01334 not applicable  APWU01335 not applicable  APWU01336 not applicable  APWU01337 not applicable  APWU01338 not applicable  APWU00208 not applicable  APWU00529 not applicable  APWU00530 not applicable  APWU00531 not applicable  APWU00532 not applicable  APWU00533 not applicable  APWU00756 not applicable  APWU00757 not applicable  APWU00758 not applicable  APWU00759 not applicable  APWU00760 not applicable  APWU00761 not applicable  APWU00762 not applicable  APWU00763 not applicable  APWU02115 010084068210229221APWU0211511151007  APWU02116 010084068210229221APWU0211611151007  APWU02117 010084068210229221APWU0211711151007  APWU02118 010084068210229221APWU0211811151007  APWU02119 010084068210229221APWU0211911151007  APWU02120 010084068210229221APWU0212011151007  APWU02121 010084068210229221APWU0212111151014  APWU02122 010084068210229221APWU0212211151014  APWU02123 010084068210229221APWU0212311151014  APWU02124 010084068210229221APWU0212411151014  APWU02125 010084068210229221APWU0212511151014  APWU02126 010084068210229221APWU0212611151014  APWU02127 010084068210229221APWU0212711151014  APWU02128 010084068210229221APWU0212811151014  APWU02129 010084068210229221APWU0212911151014  APWU02130 010084068210229221APWU0213011151014  APWU02131 010084068210229221APWU0213111151014  APWU02132 010084068210229221APWU0213211151014  APWU02133 010084068210229221APWU0213311151014  APWU02134 010084068210229221APWU0213411151014  APWU02135 010084068210229221APWU0213511151014  APWU01681 not applicable  APWU01682 not applicable  APWU01683 not applicable  APWU01924 010084068210229221APWU0192411150917  APWU01925 010084068210229221APWU0192511150917  APWU01926 010084068210229221APWU0192611150917  APWU01927 010084068210229221APWU0192711150917  APWU02104 010084068210229221APWU0210411151007  APWU01939 010084068210229221APWU0193911150917  APWU01940 010084068210229221APWU0194011150917  APWU01941 010084068210229221APWU0194111150918  APWU01942 010084068210229221APWU0194211150918  APWU01943 010084068210229221APWU0194311150918  APWU02105 010084068210229221APWU0210511151007  APWU02106 010084068210229221APWU0210611151007  APWU02107 010084068210229221APWU0210711151007  APWU02177 010084068210229221APWU0217711151014  APWU02178 010084068210229221APWU0217811151014  APWU01676 not applicable  APWU00509 not applicable  APWU00510 not applicable  APWU00511 not applicable  APWU00512 not applicable  APWU00990 not applicable  APWU00991 not applicable  APWU00992 not applicable  APWU00993 not applicable  APWU00994 not applicable  APWU00995 not applicable  APWU00996 not applicable  APWU00997 not applicable  APWU00998 not applicable  APWU00999 not applicable  APWU00714 not applicable  APWU00715 not applicable  APWU00716 not applicable  APWU00717 not applicable  APWU00718 not applicable  APWU00294 not applicable  APWU00508 not applicable  ANAU00102 not applicable  ANAU00103 not applicable  APWU02329 010084068210229221APWU0232911151103  APWU00973 010084068210229221APWU0097311150511  APWU00975 010084068210229221APWU0097511150511  APWU00976 010084068210229221APWU0097611150511  APWU00980 010084068210229221APWU0098011150511  APWU00480 not applicable  APWU00481 not applicable  APWU00482 not applicable  APWU00483 not applicable  APWU00484 not applicable  APWU00352 not applicable  APWU00972 010084068210229221APWU0097211150511  APWU00974 010084068210229221APWU0097411150511  APWU00977 010084068210229221APWU0097711150511  APWU00978 010084068210229221APWU0097811150511  APWU00979 010084068210229221APWU0097911150511  APWU00981 010084068210229221APWU0098111150511  APWU00251 not applicable  APWU00912 not applicable  APWU00513 not applicable  APWU00514 not applicable  APWU00558 not applicable  APWU00246 not applicable  APWU02109 010084068210229221APWU0210911151007  APWU01000 not applicable  APWU01001 not applicable  APWU01002 not applicable  APWU01003 not applicable  APWU01004 not applicable  APWU01005 not applicable  APWU01006 not applicable  APWU01007 not applicable  APWU01008 not applicable  APWU01009 not applicable  APWU01010 not applicable  APWU01011 not applicable  APWU01371 010084068210229221APWU0137111150624  APWU01372 010084068210229221APWU0137211150624  APWU01373 010084068210229221APWU0137311150624  APWU01374 010084068210229221APWU0137411150624  APWU01375 010084068210229221APWU0137511150624  APWU01376 010084068210229221APWU0137611150624  APWU00722 not applicable  APWU01436 not applicable  APWU01437 not applicable  APWU01438 not applicable  APWU01439 not applicable  APWU01440 not applicable  APWU01441 not applicable  APWU01442 not applicable  APWU00557 not applicable  APWU02030 010084068210229221APWU0203011150929  APWU02031 010084068210229221APWU0203111150929  APWU02072 010084068210229221APWU0207211151002  APWU02073 010084068210229221APWU0207311151002  APWU02074 010084068210229221APWU0207411151002  APWU02075 010084068210229221APWU0207511151002  APWU02159 010084068210229221APWU0215911151014  APWU02160 010084068210229221APWU0216011151014  APWU02161 010084068210229221APWU0216111151014  APWU02162 010084068210229221APWU0216211151014  APWU02332 010084068210229221APWU0233211151103  APWU01679 not applicable  APWU00806 not applicable  APWU00807 not applicable  APWU00808 not applicable  APWU00809 not applicable  APWU00810 not applicable  APWU00811 not applicable  APWU00812 not applicable  APWU00813 not applicable  APWU00814 not applicable  APWU00815 not applicable  APWU00816 not applicable  APWU00817 not applicable  APWU01805 010084068210229221APWU0180511150819  APWU01799 010084068210229221APWU0179911150819  APWU00884 not applicable  APWU00172 not applicable  APWU00173 not applicable  APWU00174 not applicable  APWU00488 not applicable  APWU00489 not applicable  APWU00490 not applicable  APWU00491 not applicable  APWU00492 not applicable  APWU00493 not applicable  APWU00494 not applicable  APWU00495 not applicable  APWU00496 not applicable  APWU00313 not applicable  APWU00914 not applicable  APWU00898 not applicable  APWU00872 not applicable  APWU00873 not applicable  APWU00874 not applicable  APWU00915 not applicable  APWU00836 not applicable  APWU00370 not applicable  APWU00371 not applicable  APWU00307 not applicable  APWU00412 not applicable  APWU00464 not applicable  APWU00465 not applicable  APWU00466 not applicable  APWU00467 not applicable  APWU00468 not applicable  APWU00469 not applicable  APWU00470 not applicable  APWU00471 not applicable  APWU00472 not applicable  APWU00473 not applicable  APWU02039 010084068210229221APWU0203911150929  APWU00319 not applicable  APWU00320 not applicable  APWU00321 not applicable  APWU00322 not applicable  APWU00323 not applicable  APWU00840 not applicable  APWU00841 not applicable  APWU00842 not applicable  APWU00843 not applicable  APWU00844 not applicable  APWU00845 not applicable  APWU00846 not applicable  APWU00847 not applicable  APWU00247 not applicable  APWU00936 not applicable  APWU00211 not applicable  APWU00212 not applicable  APWU00213 not applicable  APWU00214 not applicable  APWU01761 010084068210229221APWU0176111150819  APWU01762 010084068210229221APWU0176211150819  APWU01763 010084068210229221APWU0176311150819  APWU01764 010084068210229221APWU0176411150819  APWU01765 010084068210229221APWU0176511150819  APWU01766 010084068210229221APWU0176611150819  APWU00891 not applicable  APWU00892 not applicable  APWU00343 not applicable  APWU00933 not applicable  APWU00935 not applicable  APWU01220 not applicable  APWU01221 not applicable  APWU01222 not applicable  APWU01223 not applicable  APWU01224 not applicable  APWU01225 not applicable  APWU01226 not applicable  APWU01227 not applicable  APWU01228 not applicable  APWU01229 not applicable  APWU01230 not applicable  APWU01231 not applicable  APWU01232 not applicable  APWU01233 not applicable  APWU01234 not applicable  APWU01235 not applicable  APWU01236 not applicable  APWU01237 not applicable  APWU01238 not applicable  APWU01239 not applicable  APWU01255 not applicable  APWU01256 not applicable  APWU01257 not applicable  APWU01258 not applicable  APWU01268 not applicable  APWU01269 not applicable  APWU01270 not applicable  APWU01271 not applicable  APWU01272 not applicable  APWU01273 not applicable  APWU01274 not applicable  APWU01275 not applicable  APWU01276 not applicable  APWU01277 not applicable  APWU01278 not applicable  APWU01279 not applicable  APWU01280 not applicable  APWU01281 not applicable  APWU01282 not applicable  APWU01283 not applicable  APWU01284 not applicable  APWU01285 not applicable  APWU01286 not applicable  APWU01287 not applicable  APWU01288 not applicable  APWU01289 not applicable  APWU01206 not applicable  APWU01207 not applicable  APWU01208 not applicable  APWU01209 not applicable  APWU01210 not applicable  APWU01211 not applicable  APWU01212 not applicable  APWU01213 not applicable  APWU01214 not applicable  APWU01215 not applicable  APWU01216 not applicable  APWU01217 not applicable  APWU01218 not applicable  APWU01219 not applicable  APWU01458 not applicable  APWU01459 not applicable  APWU01460 not applicable  APWU00931 not applicable  APWU00932 not applicable  APWU00391 not applicable  APWU00450 not applicable  APWU00615 not applicable  APWU00616 not applicable  APWU00617 not applicable  APWU00618 not applicable  APWU00233 not applicable  APWU00234 not applicable  APWU00235 not applicable  APWU00236 not applicable  APWU00237 not applicable  APWU00238 not applicable  APWU01928 010084068210229221APWU0192811150917  APWU00930 not applicable  APWU00939 not applicable  APWU00940 not applicable  APWU00941 not applicable  APWU01029 not applicable  APWU01030 not applicable  APWU01031 not applicable  APWU01032 not applicable  APWU01033 not applicable  APWU01034 not applicable  APWU01035 not applicable  APWU01036 not applicable  APWU01037 not applicable  APWU01038 not applicable  APWU01039 not applicable  APWU01040 not applicable  APWU01041 not applicable  APWU01042 not applicable  APWU01043 not applicable  APWU01044 not applicable  APWU01045 not applicable  APWU01046 not applicable  APWU01047 not applicable  APWU01048 not applicable  APWU01049 not applicable  APWU01050 not applicable  APWU01051 not applicable  APWU00679 not applicable  APWU00680 not applicable  APWU00681 not applicable  APWU00682 not applicable  APWU00683 not applicable  APWU00684 not applicable  APWU00685 not applicable  APWU00686 not applicable  APWU00687 not applicable  APWU00688 not applicable  APWU00689 not applicable  APWU00690 not applicable  APWU00691 not applicable  APWU00692 not applicable  APWU00693 not applicable  APWU00694 not applicable  APWU00695 not applicable  APWU00696 not applicable  APWU00697 not applicable  APWU00698 not applicable  APWU00699 not applicable  APWU00700 not applicable  APWU00701 not applicable  APWU00702 not applicable  APWU00703 not applicable  APWU00704 not applicable  APWU00705 not applicable  APWU01661 not applicable  APWU01662 not applicable  APWU01663 not applicable  APWU01664 not applicable  APWU00199 not applicable  APWU00200 not applicable  APWU00201 not applicable  APWU00202 not applicable  APWU00203 not applicable  APWU00204 not applicable  APWU00205 not applicable  APWU00206 not applicable  APWU00207 not applicable  APWU00421 not applicable  APWU00422 not applicable  APWU00423 not applicable  APWU00424 not applicable  APWU00425 not applicable  APWU00426 not applicable  APWU00427 not applicable  APWU00428 not applicable  APWU00429 not applicable  APWU00430 not applicable  APWU00431 not applicable  APWU00432 not applicable  APWU00433 not applicable  APWU00434 not applicable  APWU00435 not applicable  APWU00436 not applicable  APWU00437 not applicable  APWU00438 not applicable  APWU00439 not applicable  APWU00440 not applicable  APWU00441 not applicable  APWU00442 not applicable  APWU00443 not applicable  APWU00444 not applicable  APWU00445 not applicable  APWU00446 not applicable  APWU00447 not applicable  APWU00367 not applicable  APWU00368 not applicable  APWU00369 not applicable  APWU01461 not applicable  APWU01462 not applicable  APWU01463 not applicable  APWU01512 not applicable  APWU00209 not applicable  APWU00374 not applicable  APWU00375 not applicable  APWU00376 not applicable  APWU00377 not applicable  APWU00378 not applicable  APWU00908 not applicable  APWU00909 not applicable  APWU00910 not applicable  APWU00911 not applicable  APWU00405 not applicable  APWU00406 not applicable  APWU00407 not applicable  APWU00408 not applicable  APWU02037 010084068210229221APWU0203711150929  APWU02038 010084068210229221APWU0203811150929  APWU02331 010084068210229221APWU0233111151103  APWU01740 not applicable  APWU00838 not applicable  APWU00546 not applicable  APWU00549 not applicable  APWU00550 not applicable  APWU00790 not applicable  APWU00518 not applicable  APWU00521 not applicable  APWU00522 not applicable  APWU00523 not applicable  APWU00359 not applicable  APWU00399 not applicable  APWU01406 not applicable  APWU01946 010084068210229221APWU0194611150918  APWU00706 not applicable  APWU01833 010084068210229221APWU0183311150819  APWU00890 not applicable  APWU01846 010084068210229221APWU0184611150831  APWU01847 010084068210229221APWU0184711150831  APWU01848 010084068210229221APWU0184811150831  APWU01849 010084068210229221APWU0184911150831  APWU00327 not applicable  APWU00328 not applicable  APWU00329 not applicable  APWU00330 not applicable  APWU00331 not applicable  APWU00332 not applicable  APWU00333 not applicable  APWU00334 not applicable  APWU00335 not applicable  APWU00336 not applicable  APWU00337 not applicable  APWU00338 not applicable  APWU00339 not applicable  APWU00340 not applicable  APWU00341 not applicable  APWU00342 not applicable  APWU01656 not applicable  APWU01657 not applicable  APWU01658 not applicable  APWU01659 not applicable  APWU00392 not applicable  APWU00393 not applicable  APWU00394 not applicable  APWU00395 not applicable  APWU00396 not applicable  APWU00419 not applicable  APWU00420 not applicable  APWU00451 not applicable  APWU00452 not applicable  APWU00459 not applicable  APWU00460 not applicable  APWU00462 not applicable  APWU00463 not applicable  APWU00314 not applicable  APWU02036 010084068210229221APWU0203611150929  APWU00792 not applicable  APWU00793 not applicable  APWU00461 not applicable  APWU00851 not applicable  APWU00852 not applicable  APWU00853 not applicable  APWU00854 not applicable  APWU00855 not applicable  APWU00856 not applicable  APWU00857 not applicable  APWU00858 not applicable  APWU00859 not applicable  APWU00860 not applicable  APWU00861 not applicable  APWU00862 not applicable  APWU00863 not applicable  APWU00239 not applicable  APWU00453 not applicable  APWU02154 010084068210229221APWU0215411151014  APWU02155 010084068210229221APWU0215511151014  APWU02156 010084068210229221APWU0215611151014  APWU00175 not applicable  APWU02353 010084068210229221APWU0235311151112  APWU02354 010084068210229221APWU0235411151112  APWU02355 010084068210229221APWU0235511151112  APWU02356 010084068210229221APWU0235611151112  APWU02357 010084068210229221APWU0235711151112  APWU02358 010084068210229221APWU0235811151112  APWU02359 010084068210229221APWU0235911151112  APWU02360 010084068210229221APWU0236011151112  APWU02361 010084068210229221APWU0236111151112  APWU01513 not applicable  APWU01514 not applicable  APWU00249 not applicable  APWU00379 not applicable  APWU00380 not applicable  APWU00381 not applicable  APWU00382 not applicable  APWU00383 not applicable  APWU00384 not applicable  APWU00385 not applicable  APWU00386 not applicable  APWU00387 not applicable  APWU00388 not applicable  APWU00389 not applicable  APWU00390 not applicable  APWU00798 not applicable  APWU00837 not applicable  APWU00220 not applicable  APWU00221 not applicable  APWU00547 not applicable  APWU00548 not applicable  APWU02255 010084068210229221APWU0225511151029  APWU02256 010084068210229221APWU0225611151029  APWU02257 010084068210229221APWU0225711151029  APWU02258 010084068210229221APWU0225811151029  APWU02259 010084068210229221APWU0225911151029  APWU02260 010084068210229221APWU0226011151029  APWU02261 010084068210229221APWU0226111151029  APWU02262 010084068210229221APWU0226211151029  APWU02263 010084068210229221APWU0226311151029  APWU02264 010084068210229221APWU0226411151029  APWU02265 010084068210229221APWU0226511151029  APWU02266 010084068210229221APWU0226611151029  APWU02267 010084068210229221APWU0226711151029  APWU02268 010084068210229221APWU0226811151029  APWU02269 010084068210229221APWU0226911151029  APWU02270 010084068210229221APWU0227011151029  APWU02271 010084068210229221APWU0227111151103  APWU02272 010084068210229221APWU0227211151103  APWU02273 010084068210229221APWU0227311151103  APWU02274 010084068210229221APWU0227411151103  APWU02275 010084068210229221APWU0227511151103  APWU02276 010084068210229221APWU0227611151103  APWU02277 010084068210229221APWU0227711151103  APWU02278 010084068210229221APWU0227811151103  APWU00228 not applicable  APWU00229 not applicable  APWU00230 not applicable  APWU00231 not applicable  APWU00232 not applicable  APWU00831 not applicable  APWU00832 not applicable  APWU02335 010084068210229221APWU0233511151103  APWU02336 010084068210229221APWU0233611151103  APWU02337 010084068210229221APWU0233711151103  APWU02338 010084068210229221APWU0233811151103  APWU02339 010084068210229221APWU0233911151103  APWU02340 010084068210229221APWU0234011151103  APWU02341 010084068210229221APWU0234111151112  APWU02342 010084068210229221APWU0234211151112  APWU02343 010084068210229221APWU0234311151112  APWU02344 010084068210229221APWU0234411151112  APWU00354 not applicable  APWU00355 not applicable  APWU00356 not applicable  APWU00357 not applicable  APWU00358 not applicable  APWU00295 not applicable  APWU00298 not applicable  APWU00299 not applicable  APWU00300 not applicable  APWU00301 not applicable  APWU00360 not applicable  APWU01293 not applicable  APWU01294 not applicable  APWU01295 not applicable  APWU00830 not applicable  APWU00215 not applicable  APWU00216 not applicable  APWU00217 not applicable  APWU00218 not applicable  APWU00219 not applicable  APWU01809 010084068210229221APWU0180911150819  APWU01810 010084068210229221APWU0181011150819  APWU01811 010084068210229221APWU0181111150819  APWU01812 010084068210229221APWU0181211150819  APWU01813 010084068210229221APWU0181311150819  APWU01650 not applicable  APWU01651 not applicable"
    }
  ]
}