  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Atlanta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73761",
      "recalling_firm": "Elekta, Inc.",
      "address_1": "400 Perimeter Center Ter NE Ste 50",
      "address_2": "N/A",
      "postal_code": "30346-1227",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of AK, AZ, AR, CA, CO, CT, FL, GA, HI, IL, IN, KY, LA, MA, MI, MN, MS, MO, MT, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA, WA, WV, WI, and countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Columbia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Indonesia, Iraq, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Saudi Arabia, Serbia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom",
      "recall_number": "Z-1561-2016",
      "product_description": "Radionuclide Radiation Therapy System    Product The product is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters.",
      "product_quantity": "217",
      "reason_for_recall": "The latches of the frame adapter can be locked even if the locating pins of the frame adapter is not inserted into the corresponding holes in the coordinate frame.",
      "recall_initiation_date": "20160419",
      "center_classification_date": "20160422",
      "termination_date": "20210716",
      "report_date": "20160504",
      "code_info": "Perfexion - 14753-099/6052, 12579-007/6250, 10394-LGK01/6144, 11920-C5210/6088, 17397-001/6158, 105528-001/6122, 14857-001/6028, 12236-006/6097, 12473-003/6005, 14857-001/6028, 16867-001/6166, 12668-003/6145  10208-003/6215, 10698-001/6237, 10979-002/6213, 10316-500/6210, 21061-001/6195, 20364-001/6103, 11720-002/6143, 12620-002/6120, 10611-001/6081, 12846-004/6064, 11524-002/8001, 11420-001/6212, 11420-001/6228, 14983-001/6076, 11083-HOM001/6070, 10247-002/6136,10684-002/6090,11868-004/6002, 10795-K08002/6063, 20354-ENS/6160, 11792-005/6077, 11291-001/6018, 10326-003/6105, 18630-001/6031, 17322-001/6121, 30000656-002/6227, 12629-003/6026, 10924-001/6089, 11154-003/6050, 20438-003/6013  17522-001/6214, 17525-002/6038, 11414-LGK002/6205, 11940-LGK002/6181, 10514-LGK002/6183, 11639-LGK001/6042, 11390-LGK001/6109, 10416-LGK001/6125, 11670-LGK001/6133, 11990-LGK001/6118, 10648-LGK001/6098,  15104-LGK001/6080, 10654-LGK001/6096, 10959-LGK001/6132, 10997-LGK001/6087, 15748-LGK001/6106, 11149-LGK002/6206, 11986-LGK002/6218,  10929-LGK002/6175, 10879-LGK002/6234, 11434-LGK002/6255, 11991-LGK002/6198, 13480-LGK002/6148, 10425-LGK002/6254, 10583-LGK002/6161,  11563-LGK002/6162, 10103-LGK02/6149, 10541-LGK002/6142, 10848-LGK001/6108, 11433-LGK001/6116, 14900-LGK001/6058, 15045-LGK001/6091  11784-LGK001/6055, 30000732-001/6223, 40017-002/6170, 12085-006/6163,   12109-002/6169, 12959-002/6033, 10209-001/6046, 10960-005/6193, 0002718-001/6113, 30001751-WOJ701/6068, 12451-001/6019, 14878-003/6067, 105618-001/6110, 19634-001/6155, 16647-001/6187, 14891-001/6104, 30001527-001/6231, 12028-LGK002/6217, 11586-001/6157, 10728-001/6062, 11999-001/6126, 15416-001/6086, 10459-001/6123, 10158-001/6177, 10234-002/6111, 40023-001/6178, 10444-001/6085, 17033-003/6043, 11530-001/6101, 11553-001/6024, 10779-07/6069, 13465-CHUV03/6094, 10424-001/6182, 13370-002/6075, 12214-003/6007, 30000655-001/6199, 10152-001/6159, 11550-001/6065, 10118-001/6230, 12766-001/6112, 13160-001/6039, 11042-001/6003,   12270-001/6017, 18062-001/6146, 11710-004/6141, 11820-006/8006, 11570-004/6220, 11127-009/6219, 12164-004/6221, 10434-004/6244, 10542-012/6238,  13826-002/6239, 10567-041/6191, 98300-003/6140, 10567-037/6119, 14569-001/6135, 10643-005/6117, 11798-009/6060, 17357-001/6130, 11813-003/6034,  10356-007/6079, 10567-045/6232, 10567-032/6074, 10567-030/6036, 10567-025/6029, 11293-004/6004, 12269-002/6082, 11772-004/6048, 10164-004/6009,   10349-003/6066, 10567-026/6011, 10560-003/6127, 11073-008/6020, 10567-036/6102, 14236-001/6114, 10567-035/6092, 10379-002/6061, 11030-006/6030,  13050-001/6037, 10909-009/6010, 11714-003/6072, 12985-003/6025, 14030-002/6015, 10567-031/6054, 10064-008/6021, 13558-001/6099, 10567-038/6107, 11939-003/6008, 10591-002/6083, 10567-024/6014, 11848-003/6095, 11624-003/6078, 10383-007/6100, 11395-010/6032, 11198-005/6124, 10982-003/6053,   11256-001/6012, 10905-004/6016, 10387-011/6093, 11138-004/6115, 10524-003/6073, 10908-003/6045, 10762-005/6057, 10193-007/6056, 10673-016/6006,  11007-005/6035, 11791-006/6022, 11377-003/6186, 12752-004/6242, 11039-006/6188, 10597-002/6167, 17539-001/6185, 14153-001/6197, 30000118-001/6240, 10567-039/6180, 15053-001/6179, 11731-005/6200, 11090-005/6165,  11077-007/6207, 30000171-002/6172, 10760-006/6173, 11733-001/6192, 10817-009/6204, 13514-001/6211, 10877-GK01/6151, 10436-009/6150, 11396-003/6176, 10917-004/6128, 11373-002/6184, 10567-042/6129, 10567-043/6203,  12614-003/6139, 10297-006/6171, 11347-003/6134    LEKSELL GAMMA KNIFE ICON - 11868-008/6131, 11290-ENS1/6174, 12027-EZT-G1/6041, 10444-LGK004/8003, 11352-001/6044, 10622-005/6194, 10886-013/6027, 11201-004/6047, 11834-003/6222"
    }
  ]
}