  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mentor",
      "address_1": "5960 Heisley Rd",
      "reason_for_recall": "STERIS has identified that the control board software in select AMSCO 400 and AMSCO C units will interrupt and cancel a processing cycle should the selected sterilizer shutdown time coincide with a processing cycle.  This may result in a procedure delay.",
      "address_2": "",
      "product_quantity": "701 units",
      "code_info": "AMSCO 400 Units: Serial #'s: 030051501 - 033641440; AMSCO C Units: Serial #'s: R020051501 - R023641403.",
      "center_classification_date": "20150430",
      "distribution_pattern": "Worldwide Distribution -- US, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, I, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY; and, the countries of Afghanistan, Canada, Japan, Kuwait, Mexico, Republic of Korea, & United Arab Emirates.",
      "state": "OH",
      "product_description": "AMSCO C and AMSCO 400 Steam Sterilizers, STERIS Corporation.    AMSCO C Small Sterilizer, AMSCO 400 Small Sterilizer, AMSCO 400 Medium Sterilizer.    Designed for sterilization of heat and moisture-stable materials used in healthcare facilities.",
      "report_date": "20150506",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Steris Corporation",
      "recall_number": "Z-1561-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70938",
      "termination_date": "20160120",
      "more_code_info": "",
      "recall_initiation_date": "20150219",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}