  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mentor",
      "address_1": "5960 Heisley Rd",
      "reason_for_recall": "Should the Caviwave Pro Ultrasonic Cleaning System experience a temporary interruption in electrical power during a processing cycle, the circulation pump may stop functioning.  In addition, it is possible that the ultrasonic generator may stop working.  The machine does not provide an alert notifying the user of this device malfunction.",
      "address_2": "",
      "product_quantity": "491 units",
      "code_info": "All units distributed between 12/20/2010 and 02/18/2015",
      "center_classification_date": "20150430",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH,NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI & WV and the countries of: Australia, Brazil, Canada, France, Mexico, Saudi Arabia, Spain, Taiwan & United Arab Emirates.",
      "state": "OH",
      "product_description": "Caviwave Pro Ultrasonic Cleaning System,     Product Usage:  To wash (remove tissue, blood and other contaminants) or dry laparoscopic and general surgical instruments prior to final disinfection and sterilization as required.",
      "report_date": "20150506",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Steris Corporation",
      "recall_number": "Z-1559-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70899",
      "termination_date": "20151021",
      "more_code_info": "",
      "recall_initiation_date": "20150219",
      "postal_code": "44060-1834",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}