      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73838",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide Distribution : NC, NH, IL, MA, WA, PA, MD, MN, TX, NY, SC, OH, CA, KY, MT, IA",
      "recall_number": "Z-1558-2016",
      "product_description": "SOMATOM Force, Computed Tomography x-ray system     intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission.",
      "product_quantity": "37 CTsystems",
      "reason_for_recall": "The Neonate Head protocol with the pediatric kernel Hp38 could result in artefacts and possibly lead to a misdiagnosis (i.e. either non-existing blood or liquid is mimicked in the images or actual existing blood or liquid is not depicted as expected).  There is also a risk of a potential misdiagnosis in using this protocol for surgery planning.  Investigations revealed an incorrect parameterization of the reconstruction algorithm applied for the pediatric head kernel as cause of the problem.",
      "recall_initiation_date": "20160401",
      "center_classification_date": "20160422",
      "termination_date": "20170417",
      "report_date": "20160504",
      "code_info": "SOMATOM Force Material # 10742326, serial numbers 75550, 75526, 75512, 75567, 75536, 75514, 75563, 75510, 75454, 75540, 75571, 75437, 75500, 75527, 75475, 75535, 75481, 75439, 75450, 75482, 75460, 75493, 75476, 75458, 75528, 75467, 75487, 75478, 75513, 75532, 75555, 75524, 75515, 75559, 75501, 75570, 75576, 75450, 75482, 75528, 75576"
    }
  ]
}