      "status": "Terminated",
      "city": "Baltimore",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73673",
      "recalling_firm": "Philips Visicu",
      "address_1": "217 E Redwood St Ste 1900",
      "address_2": "N/A",
      "postal_code": "21202-3315",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution -- to the stated of FL, MA, MI, MS, PA, and KS.",
      "recall_number": "Z-1557-2016",
      "product_description": "eCareCoordinator allows the customer to schedule patient tasks (i.e., take weight measurement, take blood pressure).    For medical use by professional medical staff.",
      "product_quantity": "9",
      "reason_for_recall": "The adherence check generates a software error.  Two issues occur as a result of the software error: 1) No adherence flag is generated for that day; 2) No task is generated for the date 14 days later.   An error message is generated and forwarded to operations.   Subsequently, for days in which no task was generated, patient will not get a reminder on the tablet to take a measurement.    If the pa",
      "recall_initiation_date": "20160310",
      "center_classification_date": "20160422",
      "termination_date": "20171027",
      "report_date": "20160504",
      "code_info": "453564506091 eCareCoordinator 1.3"
    }
  ]
}