  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70916",
      "recalling_firm": "Philips Medical Systems, Inc.",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - Nationwide Distribution and the countries of Australia, Austria, Argentina, Canada, China, Chile, Colombia Czech Republic, Great Britain, Germany, Egypt, Finland, France, India, Ireland, Japan, Malaysia, Macedonia, Mexico, Netherlands, New Zealand, Nepal, Peru, Poland, Russia, Saudi Arabia Singapore, Switzerland, Sweden, Spain, United Arab Emirates and United Kingdom.",
      "recall_number": "Z-1554-2015",
      "product_description": "Philips Healthcare DigitalDiagnost System X-Ray",
      "product_quantity": "54 DigitalDiagnost",
      "reason_for_recall": "The system is designed to emit a beep upon termination of an exposure.  However, if the system has been powered on for over 12 hours, the system will no longer emit this signal.  This is a failure to comply with 21CFR 1020.31(j).",
      "recall_initiation_date": "20150209",
      "center_classification_date": "20150521",
      "termination_date": "20160614",
      "report_date": "20150527",
      "code_info": "Digital Diagnost Stationary X-ray System Release 4.0.0, 4.0.1, 4.0.2"
    }
  ]
}