  "results": [
    {
      "status": "Terminated",
      "city": "Haifahaifa",
      "state": "",
      "country": "Israel",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87818",
      "recalling_firm": "Argo Medical Technologies Ltd",
      "address_1": "Matambuilding 30p.O.Box 15054",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1553-2021",
      "product_description": "ReWalk Personal 6.0, ReWalk Personal 5.0 and ReWalk R Systems orthotically fits to the lower limbs and part of the upper body and is intended to enable individuals with spinal cord injury at levels T7 to L5 to perform ambulatory functions.",
      "product_quantity": "350 units",
      "reason_for_recall": "ReWalk Robotics received two complaints (one in the US; one in the Germany), which alleged that the ReWalk device waistpack (which is the location of the Li-Ion battery packs) caught fire while being charged.",
      "recall_initiation_date": "20200228",
      "center_classification_date": "20210511",
      "termination_date": "20211110",
      "report_date": "20210519",
      "code_info": "All Serial numbers",
      "more_code_info": ""
    }
  ]
}