  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lake Forest",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68074",
      "recalling_firm": "Tenex Health Inc",
      "address_1": "26902 Vista Terrance",
      "address_2": "N/A",
      "postal_code": "92630",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (Nationwide) and Internationally to Panama and Singapore.",
      "recall_number": "Z-1552-2014",
      "product_description": "TX1 Tissue Removal System.  TX1 Procedure Pack P/N 554-1003-001 (outer box), TX1 MicroTip P/N 554-1002-001 (inner tray). For use with the TX System Console as an ultrasonic surgical aspirator of soft tissue.",
      "product_quantity": "9,927 units",
      "reason_for_recall": "The sterile barrier in the packaging may be compromised due to cracks in the thermoform tray that may occur during transport of the product in extreme cold weather conditions, in conjunction with mechanical impact.",
      "recall_initiation_date": "20140414",
      "center_classification_date": "20140506",
      "termination_date": "20140716",
      "report_date": "20140514",
      "code_info": "Lot Numbers:   00614-03, 06513-02, 24913-01,  00713-06, 06514-02, 25313-01,  00814-02, 06914-01, 25513-02,  01114-02, 07213-04, 25613-01,  01514-03, 07713-18, 26113-06,  01714-04, 08113-04, 26313-04,  01813-02, 08413-13, 26713-04,  02314-08, 09313-05, 26813-01,  02413-04, 10213-07, 27013-10,  02714-02, 10613-01, 27613-02,  02914-01, 11213-07, 28013-05,  03014-02, 11613-06, 28313-02,  03113-02, 12113-02, 29513-03,  03114-02, 12713-03, 29713-01,  03414-03, 13413-01, 30913-07,  03514-05, 14413-05, 31113-08,  03714-03, 15013-04, 31513-02,  03814-06, 15413-02, 31713-05,  04114-02, 15413-03, 32213-01,  04213-01, 16513-04, 32413-01,  04214-05, 17013-02, 32613-01,  04414-03, 17013-09, 32913-07,  04514-02, 17613-08, 33613-09,  04814-07, 17913-16, 33813-05,  05014-02, 18413-05, 34013-03,  05113-01, 18413-06, 34313-01,  05113-02, 18613-03, 34413-04,  05114-09, 19213-03, 34513-04,  05514-01, 19713-04, 34713-08,  05614-04, 20713-06, 35113-01,  05713-03, 21813-01, 35213-03,  05714-10, 23913-01, 36013-01,  06313-05, 24113-04, 36113-02,  06314-04, 24213-04, and 36413-05."
    }
  ]
}