  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Rockaway",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92088",
      "recalling_firm": "Linkbio Corp.",
      "address_1": "101 Round Hill Dr Ste 7",
      "address_2": "",
      "postal_code": "07866-1214",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "NY, NV Foreign:  Bulgaria, France, Italy, Norway, Poland, Sweden, Switzerland, Slovakia, Spain, United Kingdom, Germany, Austria, Argentina, Australia, China, India, Iraq, Jordan, Colombia, Mexico, Saudi Arabia, Ukraine, Vietnam, and Belarus",
      "recall_number": "Z-1550-2023",
      "product_description": "LINK SLED Knee System  POREX ENDO MODEL SLED TIBIAL COMP 11 MM X 55 MM\t  Item Number: 15-2230/11",
      "product_quantity": "3 units",
      "reason_for_recall": "Additional Label Warnings: Amendment / reinforcement of the Surgical Technique and future product training on the updated technique, due to early aseptic loosening or increased wear of the implant caused by suboptimal positioning of the implant or inadequate cementing technique leading to an unsatisfactory surgical result, and in the worst case, intervention or revision earlier than expected",
      "recall_initiation_date": "20230321",
      "center_classification_date": "20230504",
      "report_date": "20230510",
      "code_info": "UDI-DI: 04026575429219   All lot numbers within the labeled expiration date",
      "more_code_info": ""
    }
  ]
}