      "status": "Terminated",
      "city": "Laurel",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68067",
      "recalling_firm": "GE Healthcare",
      "address_1": "8880 Gorman Rd",
      "address_2": "N/A",
      "postal_code": "20723-5800",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-1550-2014",
      "product_description": "Panda¿ Wall-Mount Infant Warmers, integrated with Nellcor SpO2.    Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).",
      "product_quantity": "15 units",
      "reason_for_recall": "Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.",
      "recall_initiation_date": "20140402",
      "center_classification_date": "20140506",
      "termination_date": "20160511",
      "report_date": "20140514",
      "code_info": "HDJS52856  HDJS52857  HDJS52858  HDJS52859  HDJS52860  HDJS52861  HDJS52862  HDJS52863  HDJS52864  HDJS52865  HDJS52800  HDJS52801  HDJS52802  HDJS52803  HDJS52804"
    }
  ]
}