{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Biomet has initiated this action following an investigation which identified that the slot on the resection guide is offset in the wrong direction.  When the slot on the resection guide is offset in the wrong direction, the Steinman pin that is intended to hold the guide into place may not come into contact with the bone.",
      "address_2": "",
      "product_quantity": "17",
      "code_info": "Catalog number: 010001814 and lot numbers: 318520 and 644720.",
      "center_classification_date": "20130617",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of UT, TX, AL, KY, and WI, and the countries of Australia, The Netherlands, Korea, and Canada.",
      "state": "IN",
      "product_description": "Superior Approach Resection Guide Assembly.    Indicated for primary, fracture, or revision total shoulder replacement for the relief of pain.",
      "report_date": "20130626",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Biomet, Inc.",
      "recall_number": "Z-1548-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65212",
      "termination_date": "20140609",
      "more_code_info": "",
      "recall_initiation_date": "20130515",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}