  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87811",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide",
      "recall_number": "Z-1547-2021",
      "product_description": "NM/CT 870 DR Nuclear Medicine / CT Scanners",
      "product_quantity": "242 in total (17 units US)",
      "reason_for_recall": "GE Healthcare has become aware of a potential issue on 800 family NM/CT  Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the  intended radiation exposure range of the planned scan under specific workflows. The issue occurs  only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on  the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would  expose the patient to additional X-ray radiation.",
      "recall_initiation_date": "20210207",
      "center_classification_date": "20210511",
      "report_date": "20210519",
      "code_info": "GTIN*00840682140836",
      "more_code_info": ""
    }
  ]
}