  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Laurel",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68067",
      "recalling_firm": "GE Healthcare",
      "address_1": "8880 Gorman Rd",
      "address_2": "N/A",
      "postal_code": "20723-5800",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-1547-2014",
      "product_description": "Giraffe¿ Infant Warmers integrated with Nellcor SpO2.    Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitoring feature used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by a SpO2 sensor).",
      "product_quantity": "50 units",
      "reason_for_recall": "Medical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.",
      "recall_initiation_date": "20140402",
      "center_classification_date": "20140506",
      "termination_date": "20160511",
      "report_date": "20140514",
      "code_info": "HDJM50062  HDJM50093  HDJM50470  HDJM50654  HDJM51812  HDJS55099  HDJS53396  HDJS53397  HDJS53398  HDJT50393  HDJS55036  HDJS55037  HDJS55038  HDJT50318  HDJS52165  HDJS52166  HDJS52119  HDJS52120  HDJR57563  HDJR57564  HDJR57565  HDJR57566  HDJR57567  HDJR57568  HDJR57569  HDJS50733  HDJS52092  HDJS52093  HDJS51002  HDJS51003  HDJS51004  HDJS51005  HDJS51006  HDJP50601  HDJP50606  HDJN50376  HDJS50790  HDJS50791  HDJS50792  HDJS50159  HDJS50160  HDJS51426  HDJS51427  HDJS50734  HDJS50735  HDJS52118  HDJT50534  HDJT50039  HDJR58198  HDJS50316"
    }
  ]
}