  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "56 E Bell Dr",
      "reason_for_recall": "Biomet Trauma (\"Biomet\") has initiated a recall of Retrograde Femoral Connecting Bolt, which involves Part Number 14-442021.  The connecting bolt has an undersized diameter specification that may cause an interference fit with the 4mm hex driver (Part Number: 41024) near the edge of the tolerance. If the 4mm hex driver becomes stuck in the connecting bolt it cannot be taken apart and a delay in su",
      "address_2": "",
      "product_quantity": "75",
      "code_info": "Catalog number: 14-442021 and lot number:757910",
      "center_classification_date": "20130614",
      "distribution_pattern": "Worldwide Distribution - USA Nationwide: CA, TX, NY, PA, NJ, FL, OH, MI, CO, SD, IN, MT, GA, and WI. and the countries of Netherlands, Costa Rica, Japan, and San Juan.",
      "state": "IN",
      "product_description": "Biomet Trauma Phoenix Retrograde Femoral Connecting Bolt Trauma Fixation Systems.    Product Usage:  The Phoenix Retrograde Femoral Nail is indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease and the fixation of femurs that have been surgically prepared (osteotomy) for correction of deformity.",
      "report_date": "20130626",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Biomet, Inc.",
      "recall_number": "Z-1547-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "65210",
      "termination_date": "20150506",
      "more_code_info": "",
      "recall_initiation_date": "20130429",
      "postal_code": "46582-6989",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}