{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87811",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide",
      "recall_number": "Z-1546-2021",
      "product_description": "NM/CT 860 Nuclear Medicine / CT Scanners",
      "product_quantity": "242 in total (17 units US)",
      "reason_for_recall": "GE Healthcare has become aware of a potential issue on 800 family NM/CT  Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the  intended radiation exposure range of the planned scan under specific workflows. The issue occurs  only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on  the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would  expose the patient to additional X-ray radiation.",
      "recall_initiation_date": "20210207",
      "center_classification_date": "20210511",
      "report_date": "20210519",
      "code_info": "GTIN*00840682140751",
      "more_code_info": ""
    }
  ]
}