  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Laval",
      "state": "",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71061",
      "recalling_firm": "Pega Medical Inc.",
      "address_1": "1111 Autoroute Chomedy",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide distribution.  US in Delaware, Florida, Nebraska, and New York;   Australia, Brazil, and United Kingdom.",
      "recall_number": "Z-1546-2015",
      "product_description": "Female Driver, Part number FDr101, instrument used for the insertion of the Fassier-Duval IM Telescopic System.  For orthopedic procedures.      This instrument is included in the Fassier-Duval IM Telescopic System instrument tray. This instrument is marked with the catalogue number and lot number.",
      "product_quantity": "9 units (5 units in the US and 4 units international)",
      "reason_for_recall": "The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.",
      "recall_initiation_date": "20150410",
      "center_classification_date": "20150429",
      "termination_date": "20150629",
      "report_date": "20150506",
      "code_info": "Lot 290609-01",
      "more_code_info": ""
    }
  ]
}