      "status": "Terminated",
      "city": "Sparks",
      "state": "MD",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79769",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "address_2": "BD Diagnostic Systems",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to accounts in 8 states: CA, FL, NJ, NY, OR, PA, VA, and WI. International distribution to Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Europe, Hong Kong, Japan, Singapore, Taiwan, and Thailand.",
      "recall_number": "Z-1545-2018",
      "product_description": "BD MAX DNA MMK Lab Use, catalog no. 442828",
      "product_quantity": "N/A",
      "reason_for_recall": "The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.",
      "recall_initiation_date": "20180302",
      "center_classification_date": "20180428",
      "termination_date": "20180702",
      "report_date": "20180509",
      "code_info": "All since 08/20/2014"
    }
  ]
}