  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irvine",
      "address_1": "18 Technology Dr Ste 133",
      "reason_for_recall": "Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.",
      "address_2": "",
      "product_quantity": "15",
      "code_info": "Devices distributed between 01/15/2021 and 03/14/2021.  Sale Order #/Serial #:  1599930872/ 12335834, 13617863, 13447266, 13089105, 12333728, 13616451, 13618575, 12729374;  82943/ 4122252, 4122320;  01112101A/ 12638236, 13813590, 12639090, 13744738, 9951014",
      "center_classification_date": "20210510",
      "distribution_pattern": "California",
      "state": "CA",
      "product_description": "Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part",
      "report_date": "20210519",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Infusion Pump Repair",
      "recall_number": "Z-1544-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "87657",
      "more_code_info": "",
      "recall_initiation_date": "20210326",
      "postal_code": "92618-2311",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}