  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68045",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution",
      "recall_number": "Z-1544-2014",
      "product_description": "12-Instrument Sterilization Tray;    Product Usage:  The 12-Instrument Tray is used to hold and protect surgical instruments during the sterilization process. Specifically, this tray is used for the Conquest Manual Instrument line. The tray consists of an interlocking tray and lid which are perforated to allow for the passage of sterilizing agent from outside the tray to the inside of the tray where the devices are held. The tray features silicon device holders that secure devices during the sterilization process.",
      "product_quantity": "2591 trays",
      "reason_for_recall": "Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.",
      "recall_initiation_date": "20140416",
      "center_classification_date": "20140502",
      "termination_date": "20140818",
      "report_date": "20140514",
      "code_info": "Model number: 0242000012;  Lot/serial numbers:  021464-2, 39824-1-1, 4014, 643893-1-4, 27419, 42901-1-1, 35728-2-1,   48639-1-8, 021464-3, 40288-1-1, 43020, 43893-1-6, 27701, 43104-1-1,   36175-1-1, 48956-1-1, 23027, 40629-3-,1, 43622, 43893-1-7, 29129,   43270-1-1, 36421-1-1, 48956-1-1A, 23398, 40721-1-1, 50476,   43893-1-8, 29670, 43523-1-1, 36424-1-1, 51304-1-1, 23542,   41231-4-1, 54990, 43893-1-9, 30250, 43523-1-1A, 37251-1-1,   51304-1-1A, 23846, 41478-1-1, 7J1, 48639-1-2, 30919, 43674-1-1,   37392-1-1, 52123-1-1, 24073, 41745-1-1, 123, 48639-1-3, 3218.,  43893-1-1, 37794-1-1, 52956-1-1, 24511, 42086-1-1, 28646-1-1,   48639-1-4, 33226, 43893-1-10, 38867-1-1, 54346-1-1, 26445,   42306-1-1, 29005-1-1, 48639-1-5, 33961, 43893-1-1, 1SS15327, 26843,   42526-1-1, 30384-1-1, 48639-1-7, 35933, 43893-1-12, 37991, 36610,   43893-1-13, 39219, 43893-1-2, 43893-1-14."
    }
  ]
}