  "results": [
    {
      "status": "Ongoing",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "93771",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-1543-2024",
      "product_description": "Smiths Medical Medfusion Guide Barrel Clamp, Item number G6000716",
      "product_quantity": "1842 guide barrel clamps",
      "reason_for_recall": "Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.",
      "recall_initiation_date": "20231219",
      "center_classification_date": "20240415",
      "report_date": "20240424",
      "code_info": "Lot Numbers:  P0407365, P0486670, P0561740",
      "more_code_info": ""
    }
  ]
}