  "results": [
    {
      "country": "United States",
      "city": "Andover",
      "address_1": "3000 Minuteman Road",
      "reason_for_recall": "When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator  switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when  component supplie",
      "address_2": "",
      "product_quantity": "11 - US",
      "code_info": "",
      "center_classification_date": "20120613",
      "distribution_pattern": "Nationwide Distribution",
      "state": "MA",
      "product_description": "Philips Healthcare Practix Convenio Mobile X-ray system    Worldwide Distribution -- Nationwide",
      "report_date": "20120620",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Healthcare Inc.",
      "recall_number": "Z-1538-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "61830",
      "termination_date": "20170118",
      "more_code_info": "",
      "recall_initiation_date": "20111019",
      "postal_code": "01810",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}