  "results": [
    {
      "status": "Terminated",
      "city": "Hercules",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68016",
      "recalling_firm": "Bio-Rad Laboratories, Inc.",
      "address_1": "4000 Alfred Nobel Dr",
      "address_2": "N/A",
      "postal_code": "94547-1803",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distrubution ; USA INCLUDING CA, NY, GA, NC, AL, NJ, CT, WA, NY, LA, Il, MA and Internationally  to China,  Australia, Canada, France, Great Britain, Hong Kong, Thailand, Singapore,and  Brazil.",
      "recall_number": "Z-1537-2014",
      "product_description": "Variant II beta-thalassemia Short Program Model Numbers 270-2103 (250 tests) and 270-2154 (500 tests), Bio-Rad Laboratories, Inc.  in vitro diagnostic.  The Variant II Hemoglobin testing system uses the principals of high performance liquid chromatography (HPLC) for the separation and determination of normal and abnormal hemoglobin.",
      "product_quantity": "457 kits",
      "reason_for_recall": "There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affected lots are used.",
      "recall_initiation_date": "20140408",
      "center_classification_date": "20140502",
      "termination_date": "20140530",
      "report_date": "20140514",
      "code_info": "Catalog number: 270-2103 (250 tests;  Reorder pack lot number: 64005302,  Expiration: 4/2015;  Catalog number: 270-2154 (500 tests)  Reorder pack lot number: 64005207, Expiration:  6/2015; 64004750, Expiration: 11/2015"
    }
  ]
}