  "results": [
    {
      "country": "United States",
      "city": "Harleysville",
      "address_1": "1499 Delp Dr",
      "reason_for_recall": "The 9.6F port kits were packaged with the incorrect valved peelable introducer. The label states the kit contains a 10F valved peelable introducer. The kit is packaged with a 9F valved peelable introducer. The port lumen will not fit through the introducer during the insertion procedure.",
      "address_2": "",
      "product_quantity": "71 units",
      "code_info": "MRCTI96801  Lot# MHWQ060 S2 , UDI# 884908031232",
      "center_classification_date": "20170322",
      "distribution_pattern": "Worldwide Distribution to PR and Panama",
      "state": "PA",
      "product_description": "9.6F Plastic Dignity Mid-Sized CT Port w/Pre-Attached Silicone Catheter  The CT Power Injectable Implantable Infusion Ports are indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Power Injectable Implantable Infusion Port device is indicated for power injection of contrast media",
      "report_date": "20170329",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medical Components, Inc dba MedComp",
      "recall_number": "Z-1536-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76625",
      "termination_date": "20180202",
      "more_code_info": "",
      "recall_initiation_date": "20170223",
      "postal_code": "19438-2936",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}