  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Durham",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84993",
      "recalling_firm": "bioMerieux, Inc.",
      "address_1": "100 Rodolphe St",
      "address_2": "N/A",
      "postal_code": "27712-9402",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution in the countries of Germany and the Netherlands.",
      "recall_number": "Z-1534-2020",
      "product_description": "VITEK 2 card AST-N371 - Product Usage: The VITEK 2 is an automated system consisting of instruments, software, and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.",
      "product_quantity": "4,130 kits (82,600 cards)",
      "reason_for_recall": "The top seal of some of the pouches was compromised which can allow moisture to enter that can impact some antibiotics on the card.",
      "recall_initiation_date": "20200205",
      "center_classification_date": "20200320",
      "termination_date": "20221028",
      "report_date": "20200401",
      "code_info": "REF: 422024  Lot Number: 0210932204",
      "more_code_info": ""
    }
  ]
}