{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68141",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- USA, except ND, RI and WY, and including DC, Guam, and Puerto Rico; and, the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA,AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, CANADA, CAYMAN ISLANDS, CHINA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, FRENCH POLYNESIA, GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KOREA, KUWAIT, LEBANON, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NORWAY, OMAN, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, T UREY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, UZBEKISTAN, and VENEZUELA.",
      "recall_number": "Z-1533-2014",
      "product_description": "GE Healthcare Discovery MR450, Discovery MR750, Optima MR450w, Optima MR450w equipped with the GEM option, Discovery MR750w, and Discovery MR750w equipped with the GEM option products.    Affected devices are whole body magnetic resonance scanners indicated for use as a diagnostic imaging devices.",
      "product_quantity": "1,548 units (704 US, 844 OUS).",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue with certain GE MRI Products.  Patient egress from the system can be delayed or impeded if the cradle release handle and/or cradle release block mechanism malfunctions or becomes misaligned. This could make the safe removal of the patient from the system difficult.",
      "recall_initiation_date": "20140423",
      "center_classification_date": "20140501",
      "termination_date": "20150121",
      "report_date": "20140507",
      "code_info": "Mfg Lot or Serial # SystemID  00001250FMM026 919684DMPMR1  00000301309MR1 6628  00000000HM0426 6947  00000305379MR0 7462  00001018XF503B 205592750  00001227FMM00T 201445750W  00000297006MR9 202476MR1  00000296307MR2 202476MR2  00000297727MR0 202476MR3  00000300562MR6 202476MR4  00000000UA0153 202966MR3TW  00000305414MR5 203739MR450W  00000301842MR1 203863MR450W  00001203FMMOUJ 205750MR5  00000298836MR8 205838MR4  00000298249MR4 205949MR450W  00000301841MR3 206292SR450W  00000000UA0026 206292SR750W  00000293338MR0 206CH750MR  00000295024MR4 207945MR750  00001250FMM033 207OA450WMR  00000305461MR6 208882MR450W  00000299644MR5 209225VC450  00001026XF5173 209225VC750  00001105XF502C 209SJMR  00000000UA0143 210617HUBMR  00000300706MR9 210MSMTMR  00000000UA0052 21277MR1  00000301045MR1 212989MR1  00000305990MR4 212GE450W  00000301820MR7 214266PCBCMR1  00000302494MR0 214302PDWMR1  00000300930MR5 214590MWMRE3  00001250FMM02W 214645UT3TMR1  00000030251MR2 215ARIA450W  00000304079MR7 217223BHMR  00000303322MR2 217258SBMR1  00000294098MR9 217383CMR5  00000291464MR6 217528MR1  00000000HM0640 217532DVMR  00000299560MR3 217732LMR  00000292843MR0 217788MR3  00001250FMM02E 217876DMH3T  00000307113MR1 217876FMR  00000303450MR1 218828MR  00000000UA0203 225763MR2  00000306130MR6 229228450W  00000123FMM0K4 239VA450MR  00000296369MR2 240215MR  00000302363MR7 240313MR  00000301471MR9 240683MR  00000292634MR3 2486613TMR  00000298535MR6 252436MPMR  00000299171MR9 252482MR450  00001250FMM02T 252NDMR450W  00000296070MR6 253596GHTMR  00001307FMM05K 256881MR1  00001301FMM04T 256HHMR2  00000296840MR2 256HICMR2  00000302944MR4 256ODCMR1  00000303639MR9 260426VAMR  00000301255MR6 260920MR  00000302633MR3 262656MR1  00000301867MR8 262687MR1  00000308096MR7 262785mr1  00000295802MR3 269DVMR  00000000UA0138 270798MR2  00000294057MR5 281274MR1  00000000UA0028 281274MR2  00000290050MR4 281276MR750  00000298727MR9 281348450WMR  00000296429MR4 281364MR2  00001001BD6325 281477MR2  00000304177MR9 301295MR3T2  00000295215MR8 301295MR4  00001106XF5046 301295MR5  00000301518MR7 301295MR6  00000000UA0027 3014643TMR  00000300687MR1 301496MR3T4  00001301FMM047 301570MR  00000000HM0570 301618MR2  00001234FMM0J4 302993MR  00000298675MR0 303278MR  00000000HM0583 303338MR1  00000000HM0591 303338MR2  00000303940MR1 303426MR  00000304578MR8 303427MRW  00000303646MR4 303444MR  00000000UA0029 303604MR2  00000299213MR9 303699MR2  00000294494MR0 304256MR2  00000302631MR7 304766MR4  00000302248MR0 305243MR1  00001245FMM03P 305284MR1  00000000UA0085 305461MR1  00000304493MR0 305854MR1  00000301046MR9 308398HDMR  00001301FMM04G 309344MR450W  00000296469MR0 309662JMR450W  00000305101MR8 309672MR450W  00000296046MR6 309683MNMR750  00000305519MR1 310289MR  00000303805MR6 312355DVMR  00000301924MR7 312996MR3T  00000299952MR2 313593MR450  00000000UA0091 314577CGMR3T  00000301262MR2 315SOS450W  00000305521MR7 315SOSNORTH  00000303342MR0 316685MR  00001250FMM036 317338MR5  00000298399MR7 317865MR3  00000301519MR5 317887MR2  00000301190MR5 318769MR  00000299522MR3 319268MR  00000304355MR1 321FCMR1  00001301FMM045 330263450W  00000000UA0116 330344MR750W  00000300847MR1 330363450W  00000305176MR0 330489MR450W  00000000UA0144 337289MR1  00000000UA0074 337943MR1  00000029854MR5 352867MR  00000300987MR5 360507450WCM  00000298512MR5 360582450WPA  00000000UA0056 386FHPCMR3T  00000296908MR7 386FMR3T  00000302570MR7 402228MR  00000301013MR9 404329MR  00000000UA0034 405271OUCMR  00000299491MR1 405359EMR  00000302712MR5 405486MR  00000293561MR7 405752I3T  00000304151MR4 405912MMR  00001219FMM0GV 406375HMR  00000299141MR2 406543MR450W  00000305668MR6 407PMCMR  00000000UA0068 407SMSLMR1  00000000UA0083 408356MR1  00000302916MR2 412372MR1  00000299967MR0 412623MR1  00000300564MR2 412623SDHYMR3  00000298677MR6 412647PUHMR1  00000300494MR2 412647PUHMR2  00000298960MR6 412835MR1  00000306066MR2 412942MR2  00001307FMM01X 413568MR  00000000UA0167 414647MR1  00000298889MR7 414649WB3  00000298552MR1 414805MR1  00000298591MR9 414805MR2  00000300122MR9 414805MR4  00000297624MR9 415493VMR2  00000290202MR1 415723CHMR2  00000305165MR3 415723CHMR3  00000000HM0584 415723CHMR4  00000301295MR2 415723SCHMR2  00000291489MR3 415723SHMR10  00000000UA0057 415723SHMR12  00000000UA0064 415723SHMRW  00000000UA0073 415CB3TW  00000291877MR9 415L3TMR1  00000290773MR1 415L3TMR2  00000293261MR4 415OI3TMR  00000297877MR3 417COXMR1  00000302634MR1 417STJ450  00000303356MR0 417STJ4502  00000300704MR4 419407MR1  00000304799MR0 419893MR  00001307FMM019 419ARROWMR  00001301FMM03M 419PERRYMR  00001308FMM04V 419PRODEF  00000298932MR5 419WYMR  00000300906MR5 423553MR450  00000305617MR3 423844MR1  00000298908MR5 425111SMRW  00000299520MR7 425222SMRW  00000299823MR5 425450PMR  00000300713MR5 425688MR  00000299908MR4 425750PMR  00000300588MR1 434799MR1  00000301044MR4 434MJMR1  00000293646MR6 434UVAMR  00000000UA0089 435251MR1  00000297833MR6 435716MR1  00000302571MR5 441MR450  00000303375MR0 469820BMWMR1  00000305524MR1 479452450W  00000298861MR6 479757NW450  00000000HM0679 479964450W  00000298981MR2 480425MR6  00000294665MR5 480768MR2  00000301522MR9 484503MR  00000300473MR6 501202MR1  00000305102MR6 501221LEGACY450  00000304262MR9 501225450W1  00000306028MR2 501225450W2  00000302998MR0 501661VA450  00001250FMM021 502624MR2  00000303994MR8 503220MR1  00001213FMM0LF 504842MR2  00000299851MR6 505609MR  00000300967MR7 508548MR  00000300565MR9 508693MVMR  00000304851MR9 509332MR450W  00000299321MR0 509473DMR3T  00001236FMM01X 509482450W  00000298870MR7 509546MR450W  00000000HM0629 509837MR450W  00000000HM0461 509TCOMR450W  00000307611MR4 510449MR450W  00000301926MR2 510537MR750  00000306095MR1 512816SAMR  00000302545MR9 512LRMCMR  00000305982MR1 513732MR  00000306029MR0 513CCHGTMR  00001236FMM021 513MWMR1  00001106XF509C 515263MR450  00000296171MR2 515358MR1  00000296888MR1 515643MR450  00000000HM0529 516HSSMR  00000291397MR8 517353DVMR  00000298083MR7 517450W  00000030154MR0 517HILLMR  000005148810-3 518561MR  00000296841MRO 518563MRI  00000304440MR1 518579MR  00000306027MR4 520ACPMR2  00000301380MR2 520SMICMR  00000300764MR8 530876MR450  00000294556MR6 540665MR3T  00000299966MR2 540MIF3T750  00001250FMM020 541276MR450W  00000305667MR8 541317MR  00000292431MR4 541CMRIMR  00000297134MR9 541SAGHMR  0000000UA0001A 559734VIMR1  00000304494MR8 561747MR2  00000300900MR8 561795MR2  0000000UA0090A 561844MR1  00000298580MR2 570271MR6  00000296626MR5 573556MR450  00000305820MR3 573556MRW2  00000304763MR6 573756MAMR  00000296096MR1 573884MRI  00000301028MR7 574267MR  00000303812MR2 574335MR  00000305426MR9 574546MR  00000301401MR6 575622MRW1  00000305520MR9 580548BWB  00000000UA0071 585275MR750W  00000304508MR5 585785MR750  00000304720MR6 601354MR450W  00000300038MR7 601450W  00000304017MR7 601932MR450W  00000296462MR5 602588MR2  00000294576MR4 602830MR3  00001201FMM08B 602839MR1  00000000UA0096 6033343TW  00000295228MR1 6036503T  00000304719MR8 605322MR  00000303050MR9 605725MR  00000302688MR7 607CORNELL3T  00000299266MR7 608263MR4I  00000000UA0157 608263MR5W  00000290457MR1 608263MR6  00000296418MR7 608287MR1  00001052XF5031 608374MR  00000000HM0606 608775MR1  00000000UA0159 608775MR3T  00000307535MR5 608CHILDMR1  00000290957MR0 608HERIMR  00000288408MR8 608WAISMR  00000292528MR7 608WIMRMR2  00000299172MR7 609407MR  00000290659MR2 610740MR3  00000305910MR2 610997MRW  00000303645MR6 614257MR2  00000302194MR6 614566MR1  00000000UA0038 61456MR5  00000000UA0041 614GMC3T  00000303041MR8 614RMHMR3  00000303310MR7 615327MR450  00000296517MR6 615MTMR1  00000298420MR1 6162673T750  00000000UA0117 616267IMRI  00000298334MR4 616267MR450  00000298907MR7 616391450W  00000291363MR0 616391DVMR  00000298283MR3 616494DVMR  00000301469MR3 616774MR450W  00000000UA0030 616774MR750W  00000296640MR6 617726MR1A2  00000305451MR7 618288DVMR  00000298050MR6 618288OWMR  00000299645MR2 618997VAMR  00000000UA0075 619543HDMR4  00000306094MR4 623524MR1  00000300572MR5 630933MR3  00000000HM0484 630946MR1  00000299627MR0 631723MR450  00000305203MR2 631726MR450  00001301FMM03G 631968MR  00000303690MR2 636933JMR1  00000290137MR9 650723NCMR1  00000289309MR7 650723NCMR2  00000294472MR6 650723SSMR1  00000290704MR6 650723SSMR2  00000297374MR1 650PH3T  000000290605MR5 661KRMR750  00000305484MR8 662841MR5  00000000UA0147 671FHPMR750W  00000299264MR2 678565MR2  00000000HM0345 6970K  00000295134MR1 702616RDLMR  00000300968MR5 703205MR2  00000303773MR6 703205MR3  00000295111MR9 703391MR2  00000301599MR7 703573MR2DV  00000000UA0124 703698MR4  00000299683MR3 703805MR2  00000299544MR7 703805MR3  00000299181MR8 703858MR3T  00000304439MR3 704304DVMR  00000291980MR1 704384DVMR  00000304293MR4 704384MR450  00000305047MR3 704984DVMR  00000296829MR5 704CMCLMCMR  00000000HM0518 707521MR450W  00000303856MR9 707546MR450W  00000305485MR5 707651KVMR  00000300792MR9 707DG3T  00000000UA0050 708346MR3  00000000HM0708 713441MR6W  00000296114MR2 713441MR8  00000294398MR3 713441MR9  00000302796MR8 713566SMITH1  00000303439MR4 713566SMITH2  00000304167MR0 713566SMITH3  00001301FMM040 713792AMR13  00000299406MR9 713792CMR1  00000303614MR2 715623MRW  00000294718MR2 715838MR3T  00000300866MR1 715ADCMR1  00000306011MR8 716665MR1  00000296602MR6 7168453T  00000000HM0522 717975MR2  00000000UA0012 71872MR750W  00000299199MR0 719260MR  00000299331MR9 720321MR  00001250FMM02C 720708MR  00000302193MR8 720SAHMOBMR  00000304721MR4 724981MR450W  00000298574MR5 727462MR1  00000301126MR9 727834MR  00001251FMM00Y 731541MROP2  00000304319MR7 731644MR2  00000296860MR0 732249MR  00000302997MR2 732938MR  00001245FMM04T 7347633TMRFIX  00000000HM0612 757261mr  00000299200MR6 757312MR  00000000UA0099 757827MR2  00000294953MR5 7579533T  00000000HM0683 757953450WMR  00000000UA0141 757983MR  00000303767MR8 760452MR1  00000308080MR1 765362MR1  00001234FMM0J0 765448MR450  00000304168MR8 770252MR  00000299160MR2 770918GEMR  00000302434MR6 773525MR3  00000000HM0314 773792MRI  00000300985MR9 775SIERRAMR  00000297355MR0 775WNKAMR  00000296097MR9 781340MR2  00000300724MR2 781340MR3  00000304944MR2 781HV450MR  00000303777MR7 785623MR2  00000301869MR4 801262MR3  00000303778MR5 801268MR2  00000295112MR7 801377MR3  00000301228MR3 801475MR1  00000304852MR7 801535MR1  00000299294MR9 802371MR  00000304375MR9 803329DVMR  00000298722MR0 803434MR2  00000295092MR1 803434MRI  00000303637MR3 803751MR  00000304406MR2 804330MR450W  00000301379MR4 804WAT450  00000300732MR5 805687MR1  00001301FMM04Y 805898MR  00000305412MR9 805SCMR  0000000UA0032A 808485MR2  00000000UA0086 808545MR  00001251FMM01U 808873MR750W  00001227FMM00M 808HRAWMR  00001250FMM035 808YMMR  00000297974MR8 810235DVMR  00000298901MR0 810762MR  00000305522MR5 812265MR  00000294698MR6 812485MR3  00000300929MR7 813745MR2  00000306065MR4 814234MR450W  00000000HM0600 814237MR450W  00000304374MR2 814333MR450W  00000303855MR1 815215MR  00000300795MR2 815300MR1  00000000UA0048 815395MR2  00001251FMM03A 815399MR2  00000301156MR6 815664MR450W  00000301235MR8 815937MR  00000299559MR5 815971MR2  00000000HM0609 816276MRW  00000301378MR6 816561MRW  00000299515MR7 816698CP3T  00000296014MR4 816932DVMR  00000296040MR9 817321RATCMR1  00000304850MR1 817472TSMR1  00000299634MR6 817482BKMR1  00000303415MR4 818790MR1  00001118FMM0WD 828213MRW  00001101XF50B7 828213RMRW  00001106XF5016 828456MRW  00000301094MR9 828586MRW  00000298537MR2 832522MR1  00000300039MR5 832522MR2  00001236FMM01B 843390MR  00000296226MR0 843606MPMR  00001236FMM01W 843762CCMR  00001301FMM03V 843832SMR  00000298655MR2 845348MR  00001026XF5198 845368MRW  00001102XF506B 845454MR  00000300883MR6 845483MR  00000301198MR8 845794MRW  00000301260MR6 845943MR  00001012BD66BC 847618MR1  00001236FMM01F 847872ZMR1  00000000UA0139 847872ZMR2  00000299339MR2 847SAMCMR3  00001301FMM04E 850862MR1  00001250FMM023 850RA450W1  00001233FMM06X 856247MR1  00000304827MR9 856247MR2  00000000UA0092 856247MR3  00000000UA0146 856581MR3T  00000000HM0631 856768MR  00000305452MR5 858549HDMR1  00000000UA0118 858622HDMR1  00000950BD627C 858822HDMR1  00000301752MR2 858822HDMR2  00000944BD6172 858939MR3  00001106XF50A6 858939MR4  00001051XF5039 859301MR  00001102XF502F 859744MR450  00000304801MR4 860678450W  00000298248MR6 860749MR  00000301962MR7 863688MR  00000298310MR4 864225DVMR  00001050XF5015 864255SFMR  00001232FMM01T 864560DVMR  00001101XF50B8 864725MRI  00001049XF5011 864849DVMR  00000303692MR8 865525MR2  00000000HM0636 901384MR450  00000303765MR2 901726MR450  00000919BD5B82 903525MR750  00000305464MR0 903927MR450  00001106XF50C0 904202MR10  00000298387MR2 904202MR6  00001026XF5178 904202MR7  00000304124MR1 904202MR8  00000304536MR6 904202MR9  00001106XF50CB 904215MR11  00001049XF5008 906786WB  00001250FMM03D 9075653T  00000299230MR3 907729ANMR  00000304829MR5 907IAPMR1  00000303926MR0 908277MR1  00000298656MR0 908851MR  00001052XF5053 910323VRIMR  00000293015MR4 910907WAMR1  00000304717MR2 910CHMR450W  00001301FMM04X 912576MR  00000308068MR6 912819MR  00001106XF500D 913397DICOMRW  00000298765MR9 9155693TMR  00001101XF50BC 916961450MR  00000296512MR7 916LST3TMR  00000301424MR8 916MCCMR  00000297387MR3 916SUTROSMR  00000306030MR8 918286MR  00001027XF500C 918307MR  00001213FMM0LG 918333JPMR1  00001307FMM049 918333JPMR2  00000293065MR9 918491MR1  00000299161MR0 918523MR450  00000303693MR6 918599MR  00001301FMM04K 918744SJMR1  00000306131MR4 918744SJMR2  00000000HM0595 918893TCIMR  00000304836MR0 919684BIAC6  00000301292MR9 919684CC450W1  00000903BD55CC 919684MR5  00000304138MR1 919954RMR2  00000303813MR0 919JM450W  00000299411MR9 920288MR750  00000294280MR3 920830MR450  00000302748MR9 920885MR2  00000000UA0031 928348MR  00000305431MR9 937208MR2  00001251FMM018 937WHMR  00001213FMM0LE 952456MR450  00000302914MR7 952831MR3  00000301027MR9 952993MR1  00000303823MR9 954346MR1  00000297469MR9 954772MR1  00000304016MR9 956350MRW  00000000HM0624 970874MR  00000303664MR7 970945450W  00000306098MR5 971310MR1  00000302517MR8 973579MR  00000298869MR9 973971MR2  00000304765MR1 978287MR  00000297891MR4 985ONCMR  00000304943MR4 985SAMR  00000298834MR3 ALLIANCEOPT01  00000300793MR7 BMDACCMR1  00000000UA0128 BOSTONMEDMR1  00001251FMM01F BRIGHAM750WMR1  00000300723MR4 BRUNSWICKMR  00000302492MR4 CORT450W  00000304367MR6 DOA450W  00000296821MR2 DUPONT3T  00000000UA0069 EAST750W  00000298247MR8 FITZMR  00000302690MR3 HCIMR1805  00000299320MR2 HERITAGEMR  00000298181MR9 HSS75MR450  00000294643MR2 HSSFMR450  00000293874MR4 HSSGMR750  00000300590MR7 HSSHMR450W  00000302945MR1 HSSIMR450W  00000289310MR5 HSSMR750  00000292367MR0 INSIGHTMR3228  00000296573MR9 INSIGHTMR3231  00000298959MR8 INSIGHTMR3244  00000298573MR7 JHN450W  00000300722MR6 JOI610450W  00000305175MR2 LMCPMR  00000301866MR0 MAIN450W  00000300493MR4 MKMG450W  00000290502MR4 MR129030  00000290636MR0 MR129034  00000288907MR9 MR129035  00000294913MR9 MR129036  00000296461MR7 MR129037  00000302192MR0 MR229040  00000303616MR7 MR229042  00000303691MR0 MR229043  00000303665MR4 MR229044  00001219FMM0GT MR229045  00000000UA0150 MR229046  0000000UA0067A MR229047  00000291185MR7 MSK65MR1  00000291244MR2 MSK65MR2  00000296595MR2 MSK65MR3  00000294999MR8 MSK68MR4  00000296228MR0 MSK68MR5  00001301FMM04A OPTIMA02  00000300296MR1 ORMCMR1  00000297147MR1 PENNMEDMR  00001307FMM01V PHILLYNAVYMR  00000299503MR3 QUEEN450WMR  00000298943MR2 RADNETMR1588  00000298738MR6 RADNETMR1593  00000299173MR5 RADNETMR1594  00000301470MR1 RADNETMR1603  00000000HM0278 RADNETMR1604  00000297880MR7 RADNETMR1614  0000000UA0011A RADNETMR1629  00000000UA0060 RADNETMR1631  00000000UA0053 RADNETMR1632  00000000UA0066 RADNETMR1635  00000000UA0115 RADNETMR1639XX  00000000UA0062 RADNETMR1648  00000000HM0593 RADNETMR1650  00000000UA0149 RADNETMR1651  00000000UA0158 RADNETMR1652  00000300765MR5 RADNETMR593  00000000UA0046 RADNETMR597  00000000UA0055 RADNETMR598  00000000UA0065 RADNETMR599  00000289961MR5 SINAI750MR  00000303435MR2 SINAI8MR450W  00000303430MR3 SINAI9MR450W  00000300571MR7 SLMVMR450  00000298386MR4 SMSUSS3  00000298114MR0 SPRUCE450W  00000297493MR9 STCHRISMR  00000301308MR3 STLUKESOPTMR  00000308110MR6 801IMCMR4  0000000UA0101B 702382MR750W  00000301821MR5 UCDACCMR450  00000301597MR1 UM750MR  00000000UA0166 VAWESTROX750W  00000298494MR6 VINTAGE450  00000308112MR2 703689MR  00000308069MR4 COR526967  00000306801MR2 864529MR  00000308154MR4 406293MR450W  00000000HM0715 850431450W  00001308FMM05W 510784MR450W  00000298474MR8 302744MR750  00000000HM0731 856794MR  00001245FMM040 212543MR750  00000000HM0740 614663MR  00000000HM0714 316962MR450  00000000HM0503 RADNETMR1655  00000000UA0217 415CBMRP  0000000UA0111B 713798MR1  00000000HM0753 SMSVMS4  00000000HM0757 559325CIIMR3  00000000HM0748 432333MR  00000000HM0767 803313DVMR  00000000HM0742 METHODISTMR450W  00000000HM0666 713798MR2  00000000HM0720 972863WHWMR1  00000000UA0209 671NAVYMR  00000000UA0238 2148203TMR  00000000HM0735 503732MR1  0000000UA0132B 513636MR4  00000000HM0769 918744SJMR3  00000000UA0239 602863MR  0000000HM0566A CTCA450WGEMS  00000000HM0777 601968MR450W1  00000000HM0759 205783450W  00000000HM0773 732450WMR  00000000HM0763 702616MR450W  00000000HM0750 614HMOBMR  00000000HM0758 740454MR  00001250FMM03B 609267MR  00000000HM0594 610831450W  00000000HM0745 702PKWYMR  00000000HM0765 208678MR  00000000HM0770 405270VMR2  00000000HM0775 908859MR  00000000HM0779 NNHMR450W  00000000HM0781 319272MR  00000000HM0783 630352MRI  00000000HM0791 334418450W  00000000UA0231 913322MISHMRW  00000000UA0242 MR229048  00000000UA0253 317865MR4  00000304076MR3 209491SSMR  00001309FMM023 804CHI450W  0000000HM0702A 651968MR450W  00000000HM0539 240480MR  00000000HM0620 RADNETMR610  00000000HM0682 7493  00000000HM0734 708346MR6  00000000HM0739 510784MR450WSIL  00000000HM0749 708346MR5  00000000HM0760 7688  00000000HM0771 217528MR2  00000000HM0782 520STMMR  00000000HM0792 513870MR1  00000000HM0801 WMED450W  00000000HM0802 414778MR1  00000000HM0804 717791MR1  00000000HM0806 706845MR  00000000HM0808 803434DVMR  00000000HM0814 785239MR450  00000000UA0218 708346MR4  00000000UA0233 718226MR3T  00000000UA0234 510449MR750W  00000000UA0237 720854MR  00000000UA0244 937641MR2  00000000UA0252 763795MR1  00000000UA0260 207MMC3TMR  00000000UA0261 843815750W  00000000UA0262 949760MR3  00000000UA0264 715847MR3T  00000297637MR1 650723PSYMR  00000303927MR8 202715MR2  00000308791MR3 415MB3T  00001105XF503F ASLW750  00000000HM0744 334287MR  00000000HM0829 815395MR3  00000000HM0830 RADNETMR1660  00000000HM0832 315922MR  00000000HM0833 325670ICMR  00000000HM0835 704403DVMR  00000000UA0269 954355MR750W  00000308346MR6 MSK68MR7  00000308347MR4 MSK68MR6  00000308357MR3 650780MR750  00000308534MR7   00000308591MR7   00000308749MR1   00000308899MR4   00000308903MR4   00000309079MR2   00001251FMM01J DZ1006MR01  00000000HM0495 DZ1101MR01  00001227FMM012 DZ1011MR01  00000000HM0610 DZ1486MR01  00000306373MR2 DZ1525MR01  00000305909MR4 MRHM0560  00000000UA0136 MRUA0136  00000000HM0559 MRHM0559  00000308180MR9 MRHM0774  00000000HM0670   00001308FMM01X   00000905BD566D 910272064  00001007BD646D 910272071  00001011BD633E 910272072  00003052527MR7 910272078  00000000HM0355 910272085  00001106XF50FF 910272091  00000305463MR2 910272093  00001203FMM0VW 910273043  00000291429MR9 910274032  00001117FMM0C7 910274039  00001134FMM10P 910274040  00001203FMM0W0 910274041  00001106XF50E8 910274042  00001102XF509F 910275008  00001204FMM044 910275009  00000000HM0710 83096322871  00000000HM0605 83090346059  00000000UA0248 83096536248  00000000UA0189 83092403928  00000000UA0226 83096535911  00000000HM0772 83096537610  0000000UA0121A 83104203907  00000000HM0837 83107577438  00001011BD663A 2867211  00000000UA0270   00000309051MR1   00001236FMM052   00001309FMM057   00000299229MR5 AH5824MR01  00000000UA0258 AH4300MR01  00000000HM0815 AZ4029MR01  00000300587MR3 BH1037MR01  00000301158MR2 BY4033MR01  00000000UA0047 BY4160MR01  00000294451MR0 030020MR03  00000000UA0093 140024MR01  00000000HM0729 210036MR02  00000000UA0268 130083MR05  00000000UA0251 220078MR01  00000297898MR9 2875893  00000297156MR2 2879385  00000298115MR7 2884624  00000295060MR8 MR13905YR5  00000300472MR8 MRHM0145  00000300586MR5 MRHM0158  00000300848MR9 MRHM0177  00000300474MR4 MRHM0208  00000301230MR9 MRHM0228  00000301197MR0 MRHM0264  00000299824MR3 MRHM0266  00000301584MR9 MRHM0267  00000301502MR1 MRHM0269  00000302203MR5 MRHM0323  00000303048MR3 MRHM0342  00000304353MR6 MRHM0350A  00000302990MR7 MRHM0358  00000302572MR3 MRHM0409  00000303939MR3 MRHM0430  00000304438MR5 MRHM0448  00000305403MR8 MRHM0523  00000306982MR0 MRHM0649  00000298251MR0 MRWB0558  00000000UA0184 MRUA0184  00000307449MR9 MRHM0711  00001301FMM03E MRUA0122  00000000HM0704 MR307144MR6  00000308509MR9   00000309092MR5   00001308FMM04D   00000000HM0712 MRHM0712  00000309092MR5 MRUA0272  00000294601MR0 250727MR2  00000000UA0042 403229MR1  00000300297MR9 403382MR1  00000294119MR3 403541MR1  00000300037MR9 403670MR1  00000299547MR0 403944MR1  00000295023MR6 403955MR1  00000000HM0507 416212MR450  00000291008MR1 416486MR3T  00000303049MR1 416486MR4  00000296265MR2 416595MR1  00000300986MR7 450466MR1  00000306026MR6 506544MR1  00000305624MR9 506623MR1  00000305413MR7 506648MR3  00000304471MR6 506739MR1  00000304351MR0 506857MR2  00000000HM0514 514262MR1  00000299686MR6 514281MR450  00000298951MR5 514345BMR1  00000030090MR1 514697MR450  00000300711MR9 514848MR1  00000299628MR8 519685MR2  00000291154MR3 519685MR3T  00000297850MR0 604582MR450  00000303327MR1 604875MR750  00000303821MR3 613721MR1  00000000UA0010 613737MR1  00000304945MR9 613938MR1  00000298717MR0 780532MR1  00000304855MR0 902893MR1  00000300563MR4 905522MR3T  00001002BD6387 905527MR1  00000307053MR9 ¿¿¿ 306344MR1  00000000HM0694 403777MR1  00000000HM0701 403777MR2  00000000HM0726 905522MR2  00000306911MR9 MR102094WH0  00001250FMM02U   00000297408MR7 CLCMR1  00001246FMM03C 82427120107  00000305164MR6 82427260003  00001245FMM03W 82427130071  00001245FMM04R 82427100202  00001106XF50D7 82427010056  00001101XF5019 82427020036  00000303964MR1 82427020039  00001106XF50C5 82427020042  00001106XF5055 82427040116  00000298116MR5 82427040119  00000301927MR0 82427040120  00001101XF5022 82427040121  00001101XF5016 82427040123  00001106XF5043 82427040127  00001106XF50DF 82427040132  00001106XF50B4 82427050028  00001101XF5035 82427070075  00001106XF508E 82427070079  00001106XF502D 82427090148  00001106XF504D 82427090169  00001101XF5027 82427100147  00001106XF507A 82427100149  00001106XF5125 82427120092  00001106XF512A 82427120097  00001101XF5033 82427130057  00001044XF5030 82427140074  00001106XF50C3 82427140076  00001106XF50D0 82427140087  00000300766MR3 82427150038  00001105XF502F 82427160088  00001026XF518E 82427180022  00001009BD65A1 82427210068  00001048XF500F 82427210091  00001003BD643B 82427230048  00000300963MR6 82427230051  00000302245MR6 82427230055  00000304175MR3 82427030076  00001106XF5104 82427100184  00001106XF50B6 82427270037  00001245FMM03L 82427010072  00001302FMM053 82427010073  00000000UA0200 82427120110  00000000UA0168 82427040135  00001106XF507B 82427060042  00000000UA0171 82427210109  00000306732MR9 82427040133  00000303871MR8 82427050029  00000303971MR6 82427150047  00000305117MR4 82427210081  00000305466MR5 82427040143  00000307506MR6 82427050033  00000307616MR3 82427040141  00001106XF508B 82427040128  00001302FMM04A 82427040114  00000307111MR5 82427040142  0000000UA0199A 82427040146  0000000UA0199A 82427040148  00000306612MR3 82427040153  00000306734MR5 82427140094  00000307507MR4 82427140098  00000307451MR5 82427150058  00000306731MR1 82427310057  00000307585MR0 82427310059  00000307131MR3 82427310062  00000308161MR9 82427090202  00000308162MR7 82427040137  00000308163MR5 82427120101  00000308207MR0 82427200058  00000308210MR4 82427160114  00000308358MR1 82427210116  00000308848MR1 82427040154  00000305814MR6 82427210118  00000308885MR3 82427160118  00000000UA0245   00000308885MR3   00001147FMM0EB   00001221FMM06R   00001236FMM00U   00001236FMM01N   00001301FMM03C   00001307FMM08G   00000308359MR9 82427130072  00000300966MR9 HR4199MR01  00000298847MR5 600110MR01  00000297813MR8 CS1006MR04  00000305405MR3 CS1338MR02  00000304800MR6 DK1001MR02  00000304373MR4 DK1044MR01  00000294548MR3 DK1083MR06  00000296291MR8 EG1000MR03  00000295827MR0 EG1793MR01  00000000HM0663 EG1024MR02  00001234FMM0K3 EE1012MR01  00000294713MR3 EE1019MR02  00000301912MR2 FI1015MR04  00000295485MR7 FI1052MR01  00000304492MR2 FI1052MR02  00000298431MR8 FI1120MR01  00000294397MR5 FI1128MR01  00000299453MR1 FI1149MR01  00000304580MR4 FI1175MR01  00000301520MR3 A1588529  00000296032MR6 A2032112  00000299464MR8 A4167338  00000296332MR0 A4171551  00001301FMM03H A4199398  00000296227MR2 B4498422  00000304295MR9 C2195281  00001221FMM06L C2195285  00000304538MR2 C4200438  00000302711MR7 F1680824  00000297164MR6 F4171505  00000296879MR0 M1279737  00000303376MR8 M1460210  00000301523MR7 M1656629  00000304123MR3 M171587016  00000293919MR7 M206565701  00001251FMM019 M2148250  00000000HM0634 M22006303  00000297715MR5 M226407301  00001140FMM0FV M2265406  00000298567MR9 M23084310  00000000UA0077 M2322246  00001301FMM03K M2322250  00000299341MR8 M24070301  00000301774MR6 M2427737  00000301291MR1 M24313801  00000120FMM0VC M24481307  00000297636MR3 M2634906  00000297494MR7 M2634909  00001301FMM04U M2737110  00000298242MR9 M28027100  00000305616MR5 M2848438  00000298073MR8 M2850008  00000298734MR5 M2950010  00000296141MR5 M40180117  00000294575MR6 M40330108  00001251FMM01H M40345203  00000303414MR7 M4040080  00000304077MR1 M4134645  00000292844MR8 M4152248  00000000000018 M4155416  00000298430MR0 M4159241  00000301868MR6 M4161643  00000000UA0019 M4166994  00000304555MR6 M4174545  00000295847MR8 M4182438  00000304718MR0 M4189444  00000298850MR9 M4198087  00000302544MR2 M41981124  00000300589MR9 M4200625  00000293042MR8 M4215566  00000306096MR9 M4484105  00000290985MR1 M4484966  00000297468MR1 M4490626  00000000UA0072 M45052106  00000297701MR5 M5480767  00001250FMM02X M5484837  00000297861MR7 M5484960  00000303323MR0 M5484987  00000000UA0105 M9481853  00000298385MR6 M9886845  00000000UA0070 N4192951  00000306296MR5 RE1046MR01  00000301236MR6 X106415405  00000303774MR4 X117934012  00000302996MR4 X148299404  00001112FMM084 X178710401  00000296814MR7 X2181501  00000303449MR3 X24481303  00000301125MR1 X4205001  00000301775MR3 X4211004  00000299633MR8 X4479201  00000000UA0016 X59007321  00000292928MR9 X59655523  00001220FMM01E X669977014  00000296499MR7 X687234009  00001250FMM02L X90281610  00000000HM0698 X1478401  00000000HM0721 E4167655  00000000HM0746 M4409435  00000000HM0812 LY1086MR02  00000303804MR9 RU1086MR03  00000000HM0555 RU1153MR01  00000307354MR1 RU1516MR01  00000000HM0827 RU1857MR05  00000000HM0716 RU2551MR01  00000000HM0747 RU8438MR02  00000000UA0229 RU8438MR03  00000000HM0824 X756087016  00000305675MR1   00001308FMM014   00001308FMM05R   00001309FMM05P   00000000HM0737 PF1000MR02  00000301873MR6 125995MR01  00000306010MR0 446281MR01  00000301543MR5 493519MR04  00000301913MR0 604529MR01  00000293440MR4 615010MR01  00000298665MR1 662455MR01  00000305980MR5 805534MR01  00000302543MR4 CC4400MR02  00000302991MR5 CC5340MR01  00000000UA0058 HC3068MR01  00000298676MR8 HC6474MR01  00000299261MR8 HWIL01MR01  00000301824MR9 PC0352MR02  00000000UA0163 PC3003MR01  00000000HM0491 PC4318MR02  00000288156MR3 PC7479MR08  00001236FMM013 PC7479MR10  00001301FMM051 PC8740MR02  00001117FMM0CP PC9595MR01  00000000UA0078 PC9953MR01  00000301850MR4 PKOB02MR06  00000000UA0015 PNUE08MR01  000001106XF039 RMUE01MR03  00000298332MR8 RMUE01MR04  00000000UA0114 UC2547MR01  00000292292MR0 HC1627MR04  00000000HM0658 PC4297MR02  00000000HM0733 418262MR01  00000000HM0790 220068MR01  00000000HM0700 PC1009MR01  00000000HM0795 387194MR01  00000000HM0810 HC2259MR02  00000000HM0813 HC3702MR01  00000000UA0246 HC0220MR03  00000308836MR6 123702MR02  00001117FMM0CU   00000293710MR0 H10645MR17  00000297155MR4 H1943MR12  00000300763MR0 828270036  00000000HM0685 828270037  00001307FMM04K   00000000UA0051 HU1098MR02  00000300561MR8 83027000042711  00000000UA0024 83027000049312  00000301544MR3 83027003811911  00000304336MR1 83027062157412  00000304074MR8 83027829851812  00000000UA0025 83027848548411  00000000UA0120 83027840038612  00000305046MR5 83027860027512  00000000UA0180 83027000013713  00000306902MR8 83027200005613  00000000HM0572 83027600046413  00000000HM0789 83027605456914  00000000UA0210 83027200034313  00001236FMM05Y 83049200034313  00001117FMM0C3   00001215FMMOMD   00000288909MR5 835270017  00001203FMM0V5 835270026  00000000HM0762 835270039  00000000UA0205 IE1010MR01  00000291778MR9 IL1020MR04  00000294771MR1 IL1028MR05  00000296834MR5 IL1041MR10  00000301585MR6 IL1059MR01  00000304405MR4 A126322010  00000291345MR7 A5108257  00000294545MR9 A51152177  00000302201MR9 A5127236  00000306070MR4 A5164526  00000303324MR8 A5274842  00000293613MR6 A5279314  00000000HM0601 A5295818  00001223FMM08V A5303339  00000304169MR6 A5360411  00000000000002 A5401395  00000296820MR4 A5624132  00000304354MR4 A5808528  00000305465MR7 A5808532  00000301191MR3 A5992703  00000299454MR9 B51142110  00000300591MR5 B5129914  00000000UA0063 B5129925  00000297807MR0 D5354206  00000295757MR9 F5119155  00000291184MR0 J5274809  00000000HM0743 A5312220  00000299418MR4 A5278404  00000000HM0823 A5196821  00000000HM0797   00000308748MR3   00001308FMM01C   00000295180MR4 EM0001  00000294943MR6 EM0002  00000296560MR6 EM0003  00000299147MR9 EM0004  00000299293MR1 EM0005  00000299330MR1 EM0006  00000298566MR1 EM0007  00000299545MR4 EM0009  00000299455MR6 EM0010  00000300791MR1 EM0011  00000301194MR7 EM0012  00000301577MR3 EM0013  00000301458MR6 EM0014  00000301925MR4 EM0015  00000299546MR2 EM0016  00000299407MR7 EM0017  00000301970MR0 EM0018  00000299853MR2 EM0019  00000302689MR5 EM0020  00000300734MR1 EM0021  00000302920MR4 EM0022  00000303054MR1 EM0023  00001106XF5134 EM0030  00000301586MR4 EM0101  00000302493MR2 EM0102  00000303852MR8 EM0103  00000305622MR3 EM0104  00001219FMM0H1 EM0105  00001221FMM06N EM0107  00000000HM0580 EM0108  00000302915MR4 EM0110  00000302516MR0 EM0111  00001203FMM0VD EM0119  00000303854MR4 EM0123  00000305174MR5 EM0125  00000304473MR2 EM0200  00000000UA0020 EM0201  00000000UA0018 EM0202  00000000UA0040 EM0203  00000000UA0036 EM0204  00000000UA0009 EM0205  00000000UA0106 EM0206  00000000UA0108 EM0207  0000000UA0107A EM0208  00000000UA0134 EM0209  00000000UA0094 EM0210  00000000UA0130 EM0211  00000000UA0126 EM0212  0000000UA0098A EM0213  00000000UA0135 EM0214  00000000UA0059 EM0215  00000000UA0145 EM0216  00000000UA0152 EM0217  00000000UA0133 EM0218  00000000UA0142 EM0219  00000000UA0113 EM0220  00000000UA0161 EM0221  00000000UA0109 EM0222  00000000UA0129 EM0223  00000000UA0123 EM0224  00001227FMM00C EM0225  00001250FMM02G EM0226  00001250FMM01W EM0227  00000000UA0182 EM0228  00000000UA0131 EM0230  00000000UA0140 EM0250  00000305684MR3 EM0240  00000305987MR0 EM0241  00000000HM0654 EM0112  00000000UA0196 EM0231  00000000UA0197 EM0229  00000000HM0738 EM0120  00000000HM0724 EM0115  00000000HM0761 EM0106  00000000UA0202 EM0233  00000000UA0204 EM0235  00000000UA0208 EM0234  00000000UA0213 EM0232  00000000UA0216 EM0272  00000000UA0223 EM0237  00001215FMM0ME EM0024  00000000HM0756 EM0109  00000000UA0125 EM0236  00000000UA0220 EM0271  00000295486MR5 EX0002  00000301726MR6 EX0001  00000000HM0778   00000000HM0794   00000000HM0796   00000000HM0799   00000000HM0803   00000000HM0805   00000000HM0816   00000000UA0154   00000000UA0241   00000000UA0247   00000000UA0250   00000000UA0254   00000000UA0255   00000000UA0257   00000000UA0259   00000000UA0266   00000000UA0267   00000000UA0271   00000301598MR9   00000303853MR6   00000304557MR2   00000307534MR8   00000307569MR4   00000308111MR4   00000308242MR7   00000308326MR8   00000308360MR7   00000308514MR9   00000308515MR6   00000308548MR7   00000308549MR5   00000308592MR5   00000308594MR1   00000308596MR6   00000308597MR4   00000308752MR5   00000308757MR4   00000308799MR6   00000309050MR3   00000309053MR7   00000944BD616D   00001011BD663B   00001011BD665C   00001026XF515D   00001044XF5031   00001106XF502E   00001140FMM0FL   00001203FMM0V2   00001235FMM05L   00001236FMM016   00001236FMM05B   00001236FMM05E   00001236FMM05T   00001236FMM05U   00001236FMM05V   00001308FMM00E   00001308FMM00N   00001308FMM00W   00001308FMM005   00001308FMM02J   00001308FMM02L   00001308FMM020   00001308FMM022   00001308FMM027   00001308FMM03L   00001308FMM05M   00001309FMM05Y   00001309FMM06A   00001309FMM08Y   00001203FMM0UM KA1044MR01  00000297007MR7 850270329  00000298272MR6 850270337  00000298331MR0 850270",
      "more_code_info": "341  00000299419MR2 850270347  00000301050MR1 850270351  00000300733MR3 850270367  00001101XF5039 850270370  00001026XF515A 850270377  00000300931MR3 850270385  00001204FMM04V 850270386  00000000UA0137 850270409  00000000UA0119 850270411  00000000UA0112 850270414  00001018XF5029 850270423  00000000UA0165 850270426  00000000UA0170 850270427  0000000UA0081C 850270430  00000000UA0230 850270436  00001233FMM06M 0002937140  00000301844MR7 KW1002MR01  00001236FMM01Y KW1003MR01  00000298211MR4 KW1100MR01  00000299854MR0 KW1102MR01  00001236FMM00N KW1112MR01  00000000HM0768 KW1001MR04  00000307817MR7 KW1005MR04  00001301FMM04M LB1006MR02  00000000UA0164 LB1160MR02  00000000UA0043 LB1331MR01  00000302750MR5 LB1331MR02  00000301503MR9 LB1348MR01  00000000HM0730 LB1250MR01  00000299521MR5 MQ1000MR01  00000294156MR5 2806034DVR750  00000302922MR0 MR302922MR0  00000303872MR6 MR303872MR6  00000000HM0678 MR307145MR3  00000305378MR2 MRHM0503  00000301495MR8 MRHM0562  00000307055MR4 0004169252XX  00000308183MR3 MR308183MR3  00001026XF5174 MRWB0632  00001234FMM0HT MRHM0828  00000307505MR8   00001308FMM062   00000000HM0397 MA1135MR02  00000304910MR3 AMS0424  00000292216MR9 RTD0125  00000293603MR7 RTD0127  00000303279MR4 RTD0156  00000000UA0080 RTD0191  00000301256MR4 RTD0422  00000304018MR5 TLB0229  00000304579MR6 UDN0120  00000000UA0095 UDN0121  00001106XF50E1   00001220FMM01F NC1028MR01  00000288409MR6 920279028  00001140FMM0G4 920279031  00000000HM0498 920279032  00001308FMM052 85092252519  00000297987MR0 NO1009MR02  00000295830MR4 NO1017MR03  00000304318MR9 NO1146MR02  00001106XF5121 OM1010MR01  00000296493MR0 PL1539MR01  00001106XF5132 PL1735MR02  00000306315MR3 PL1735MR03  00000296622MR4 PL1803MR01  00000303814MR8 PL2762MR01  00001105XF5047 PL2853MR01  00000305424MR4 PL2902MR01  00000000HM0613 PL1196MR01  00000000UA0190 600042MR02  00000000HM0809 PL1501MR03  00001308FMM05X PL3105MR01  00000307613MR0 PL3116MR01  00000305814MR6   00000305118MR2 MR16061YR4  00000298309MR6 QA1044MR01  00000000UA0221 QA1042MR01  00000000UA0079 RO1098MR02  00000300865MR3 600098MR01  00000304404MR7 RU1089MR05  00001233FMM06U RU1092MR02  00000301026MR1 RU1716MR01  00000301051MR9 RU1857MR03  00000305404MR6 RU2527MR01  00001101XF5070 RU2580MR02  00001251FMM032 RU3431MR01  00000296594MR5 RU3508MR01  00000300705MR1 RU3894MR01  00000301914MR8 RU3932MR01  00000303355MR2 RU6340MR01  00001105XF504C RU7762MR01  0000000UA0024A RU5599MR01  00000000HM0668 RU2526MR02  00000000HM0677 RU8282MR01  00000000HM0820 RU3637MR01  00000000HM0475 RU7620MR01  00000000UA0227 RU8438MR01  00000308901MR8 RU8541MR01  00001102XF500D   00000296613MR3 SA1011MR03  00000000HM0263 SA1011MR04  00000299965MR4 SA1012MR02  00000303776MR9 SA1031MR02  00000000UA0022 SA1033MR01  00000298266MR8 SA1052MR03  00000300882MR8 SA1058MR02  00000299951MR4 SA1114MR01  00000303644MR9 SA1175MR01  00000000HM0597 SA1255MR02  00000293984MR1 SA2180MR02  00000303666MR2 SA2281MR01  00000000HM0596 SA2315MR01  00000000HM0741 SA1014MR03  00000000HM0785 SA1037MR02  00000000HM0786 SA1004MR03  00000000HM0751 SA1026MR01  00000000HM0807 SA1188MR04  00000000HMO822 SA2143MR01  00000306645MR3 SA2304MR01  00000000HM0776 SA2308MR01  00000000HM0834 SA2385MR01  00000000HM0798 SA1253MR01  00000309114MR7   00000944BD6167   00001308FMM04K   00000000HM0615 SA1051MR02  00000000HM0723 SA1054MR03  00000000HM0732 SA1054MR04  00000000UA0127 SA1223MR01  00001251FMM00P RS4199MR01  00000306069MR6 RS4222MR01  00000000UA0023 847270026  00000301494MR1 847270027  00001301FMM04C 847270031  00001106XF5054 2939071  00000299140MR4 ZA2195MR01  00000299452MR3 CPZ43620  00000300452MR0 CPZ43623  00000301496MR6 DPN25113  00000000UA0082 DPN26332  00000302432MR0 DPN88623  00000297154MR7 MEX68909  00000000UA0045 MPX94401  00000302795MR0 NEV13613  00000301432MR1 NP739728  00000305821MR1 UPW07915  00000000HM0607 NP473629  00000293000MR6 A010MR06  00000295710MR8 A010MR07  00000000UA0039 A010MR08  00000303357MR8 A101MR01  00000304252MR0 A106MR01  00000293521MR1 AC001MR01  00000298674MR3 E001MR02  00000300298MR7 E001MR03  00000301229MR1 F009MR01  00000295538MR3 G001MR01  00000291675MR7 K002MR02  00000303993MR0 M001MR01  00000000UA0148 M001MR02  00000000UA0155 M004MR01  00000293920MR5 N002MR02  00000304828MR7 N004MR01  00000000UA0162 O009MR03  00000296547MR3 Z001MR03  00000000HM0659 A020MR04  00000308132MR0 T002MR01  00000000UA0035 120810MR02  00001251FMM00X 122002MR01  00000301293MR7 350010MR04  00000294516MR0 475290MR01  00000298944MR0 492360MR01  00000303992MR2 541010MR03  00000301971MR8 541010MR04  00000299292MR3 690120MR01  00000297386MR5 697040MR01  00000301963MR5 724010MR04  00000297175MR2 786090MR06  00000305100MR0 788000MR01  00000000UA0049 850020MR02  00000297422MR8 899010MR03  00000306909MR3 387040MR10  00000000UA0207 120810MR03  00000000HM0793 959204MR01  00000903BD55D3 853270103  00000944BD6159 853270104  00001009BD65E7 853270109  00001101XF5037 853270111  00001106XF5052 853270113  00001106XF5035 853270118  00001106XF5078 853270119  00000300907MR3 853270120  00001106XF5080 853270125  00000304853MR5 853270127  00000305636MR3 853270128  00000306097MR7 853270130  00001301FMM04J 853270135  00000000HM0766 853270137  00000000HM0825 853270140  00000000HM0680 856270062  00000000HM0621 TN1624MR01  00000000UA0110 26001MRS01  00000299646MR0 06002MRS04  00000300796MR0 06139MRS01  00000303629MR0 34006MRS04  00000000UA0102 34348MRS01  00001251FMM044 34368MRS01  00000000UA0097 AE1016MR01  00000301047MR7 AE1212MR01  00000308761MR6   00001308FMM02N   00000299340MR0 00010MRS01  00001250FMM038 00010MRS02  00000300712MR7 00040MRS01  00001233FMM070 00040MRS02  00000294967MR5 00100MRS01  00000296514MR3 00105MRS02  00000297132MR3 00109MRS01  00000298493MR8 00131MRS02  00000295709MR0 00200MRS04  00000299684MR1 00223MRS03  00000292973MR5 00242MRS02  00000298501MR8 00269MRS04  00000303638MR1 00289MRS01  00000301872MR8 00440MRS07  00000299825MR0 00487MRS03  00000296321MR3 00506MRS01  00001251FMM00V 00507MRS01  00000303822MR1 00632MRS01  00000304078MR9 00680MRS01  00000299685MR8 01305MRS01  00000000HM0660 05169MRS02  00000029804MR5 05217MRS02  00000290545MR3 10049MRS04  00000293724MR1 10050MRS03  00000303311MR5 10086MRS03  00000000HM0364 10177MRS01  00000302675MR4 10223MRS01  00000291274MR9 10307MRS03  00000301199MR6 10390MRS01  00000306126MR4 10429MRS01  00000293642MR5 10448MRS02  00000303766MR0 10580MRS01  00000000UAOO33 10639MRS01  00000303694MR4 10671MRS01  00000000UA0037 10678MRS01  00000000HM0322 10683MRS01  00000000HM0648 20483MRS01  00000293206MR9 70001MRS09  00000303309MR9 70001MRS10  00000303280MR2 70001MRS11  00000301296MR0 70043MRS05  00000298102MR5 70065MRS03  00000294332MR2 70328MRS05  00000290456MR3 70383MRS05  00000000UA0021 70431MRS04  00000000HM0567 77266MRS02  00001245FMM044 10197MRS03  00000000HM0717 77526MRS01  00000000HM0787 IE1509MR01  00000000HM0709 10462MRS02  00000000UA0265 00230MRS01  00000000HM0811 00244MRS02  00000000HMO821 10307MRS04  00000000HM0819 10537MRS01  00000308595MR8 10692MRS01  00001308FMM053   00000300689MR7 2948529MR  00000000HM0359 UZ1021MR01  00000000HM0622 UZ1010MR01  00000000HM0670 UZ1031MR01  00000304170MR4 MRHM0454  00000309113MR9   00001308FMM02X"
    }
  ]
}