{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Huntingdon Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79872",
      "recalling_firm": "Menarini Silicon Biosystems",
      "address_1": "3401 Masons Mill Rd Ste 100",
      "address_2": "N/A",
      "postal_code": "19006-3574",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and to the countries of :  Canada and Italy",
      "recall_number": "Z-1530-2018",
      "product_description": "CELLSEARCH¿ CIRCULATING Tumor Cell Kits (IVD) (Product Code 7900001)    CELLSEARCH¿ CIRCULATING Tumor Cell Kit is intended for the enumeration of circulating tumor cells (CTC) of the epithelial origin (CD45-, EpCAM+, cytokeratins 8, 18+, and/or 19+) in whole blood.",
      "product_quantity": "N/A",
      "reason_for_recall": "Unusually high number of total images/unassigned events may require additional time and effort for the operator to review all the images and may cause a delay in reporting the results.",
      "recall_initiation_date": "20180404",
      "center_classification_date": "20180426",
      "termination_date": "20190409",
      "report_date": "20180502",
      "code_info": "UDI  (01)10888385000012(17)180810(10)S038  (01)10888385000012(17)180810(10)  (01)10888385000012(17)180810(10)S038S  (01)10888385000012(17)180810(10)S038T  (01)10888385000012(17)180810(10)S038U  (01)10888385000012(17)180810(10)S038 V    Serial, Batch, Lot #  2038  S038R  S038S  S038T  S038U  S038V  S038W  (0 distributed to the field)"
    }
  ]
}