{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marcy L'Etoile",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76201",
      "recalling_firm": "BioMerieux SA",
      "address_1": "Chemin De L'Orme",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - U.S.( Nationwide); Foreign: Utd.Arab Emir., Argentina, Austria, Australia, Brazil, Canada, Switzerland, Chile, Colombia, Czech Republic, Germany, Algeria, Spain, France, United Kingdom, Greece, Hong Kong, India, Italy, Japan, South Korea, Mexico, Netherlands, Poland, Portugal, Russian, Fed. Sweden, Singapore, Thailand, Turkey, South Africa, Indonesia, Peru, Saudi Arabia, Gabon, Kazakhstan, Philippines, Ivory Coast, Venezuela, Kenya, Serbia, Tunisia, Macedonia, Lebanon, Romania, Benin, Bosnia-Herz., Oman, Uruguay,  Lithuania, Burkina Faso, Slovenia, Qatar, Kuwait, Costa Rica, Paraguay, Dominican Republic, Croatia, Panama, Congo, Dem. Rep. Congo, Honduras, Estonia, Latvia, Cambodia, Bulgaria, Vietnam, Dutch Antilles, Nicaragua, Israel, Georgia, Palestine, Egypt, Pakistan (includ.Afghanistan), Malta, Jordan Bahrain, El Salvador, Nigeria, Dutch Antilles, Iran, Belarus, Zimbabwe, Morocco, Mauretania, Anguilla, Philippines, Malaysia, Turkmenistan, Guam, Brunei, Bolivia, Guatemala, Ecuador, Belarus, Mongolia",
      "recall_number": "Z-1528-2017",
      "product_description": "ETEST Fosfomycin FM1024 Blister packaging, Product Name: ETEST Fosfomycin FM 0.064-1024.  ETEST is a quantitative technique for determining the antimicrobial susceptibility of Gram-negative and Gram-positive aerobic bacteria such as Enterobacteriaceae, Pseudomonas, Staphylococcus and Enterococcus species and fastidious bacteria, such as anaerobes, N. gonorrhoeae, S. pneumoniae, Streptococcus and Haemophilus species. The system comprises a predefined antibiotic gradient which is used to determine the Minimum Inhibitory Concentration (MIC), in ¿g/mL, of different antimicrobial agents against microorganisms as tested on agar media using overnight incubation. Current Antimicrobial Susceptibility Testing (AST) methods are based either on quantitative dilution techniques or qualitative diffusion procedures. Dilution methods are based on two-fold serial dilutions of antibiotics in broth or agar media. These methods generate the MIC value i.e. Minimum Inhibitory Concentration of a given antibiotic in ¿g/mL that will inhibit the growth of a particular bacterium under defined experimental conditions",
      "product_quantity": "unknown",
      "reason_for_recall": "Product Stability issues:  The current shelf-life claims of the ETEST products are not supported by internal testing.",
      "recall_initiation_date": "20161215",
      "center_classification_date": "20170321",
      "termination_date": "20200818",
      "report_date": "20170329",
      "code_info": "Reference no. 529140, Lot #1001390510, 1001441860, 1001609400, 1001733320, 1001761030, 1001920530, 1001966610, 1002009360, 1002084930,   1002106570, 1002158050, 1002168110, 1002395840, 1002452170, 1002500020, 1002604960, 1002992250, 1003030810, 1003127520, 1003274260, 1003322150, 1003412730, 1003535930, 1003678890, 1003800400, 1004100220, 1004177630, 1004365540, 1004410620, 1004557900, 1004703240, 1004783050, 1004838760, 1005034320, 1005097790, 1005151830; Reference no. 529148, Lot #1001368050, 1001442030, 1001566960, 1001578760, 1001708630, 1001833500, 1001894590, 1001920200, 1001955170, 1002084950, 1002172820, 1002226550, 1002410940, 1002451250, 1002501640, 1002604970, 1002964340, 1002992260, 1003121770, 1003274280, 1003491070, 1003526920, 1003626710, 1003743380, 1003946090, 1004037330, 1004273540, 1004323330, 1004557890, 1004593710, 1004696420, 1004790240, 1005031600, 1005098130, 1005151890 & 1005151920; Reference no. 529100, Lot #1001324170, 1001378120, 1001403920, 1001433710, 1001472010, 1001487340,  1001517680, 1001535750, 1001567600, 1001617880, 1001621170, 1001702730, 1001707190, 1001726300, 1001761020, 1001879240, 1001927370, 1001976740, 1001998470, 1002123080, 1002160660, 1002391350, 1002449230, 1002487920, 1002551790, 1002614480, 1002705810, 1002786510, 1002890670, 1002960380, 1002992220, 1003030800, 1003181900, 1003274090, 1003491050, 1003836110, 1004100080, 1004137920, 1004273550, 1004381380, 1004593700, 1004629890, 1004631880, 1004851860, 1005034300, 1005192380 & 1005261410 and   Reference no. 529108, Lot #1001324180, 1001370190, 1001482570, 1001527650, 1001567610, 1001634260, 1001744620, 1001829580, 1002044710, 1002090510, 1002175950, 1002223920, 1002392160, 1002449900, 1002494590, 1002565460, 1002615530, 1002706050, 1002785800, 1002960400, 1002992240, 1003099530, 1003182230, 1003322130, 1003404520, 1003491000, 1003526910, 1003580860, 1003629820, 1003876820, 1004022140, 1004177620, 1004323320, 1004576570, 1004680780, 1004941770, 1005098120, 1005212770 & 1005302500.",
      "more_code_info": ""
    }
  ]
}