{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irvine",
      "state": "CA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85069",
      "recalling_firm": "Micro Therapeutics Inc, Dba Ev3 Neurovascular",
      "address_1": "9775 Toledo Way",
      "address_2": "N/A",
      "postal_code": "92618-1811",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide,  Argentina, Australia, Austria, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Italy, Japan, Jordan, Kazakhstan, Korea, Lebanon, Mexico, Netherlands, Norway, Paraguay, Poland, Qatar, Russian Federation, Saudi Arabia, Serbia, Slovakia, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam",
      "recall_number": "Z-1527-2020",
      "product_description": "ev3 Pipeline Flex Embolization Device.  For neurological endovascular use.",
      "product_quantity": "1050",
      "reason_for_recall": "Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex embolization device has already been implanted successfully, there is no increased risk to patients due to this issue.",
      "recall_initiation_date": "20200214",
      "center_classification_date": "20200325",
      "termination_date": "20240612",
      "report_date": "20200401",
      "code_info": "Lots:  A920066  A920068  A920069  A920070  A920071  A920072  A920073  A920074  A920075  A920076  A920078  A920079  A921354  A921355  A921356  A921357  A921358  A921359  A921360  A921361  A921362  A921363  A922016  A922017  A922018  A922020  A922021  A922022  A922480  A922481  A922483  A922484  A922571  A922572  A922575  A923951  A923956  A923959  A923960  A923963  A923967  A923968  A923969  A924193  A924194  A924197  A924198  A924199  A924638  A924640  A924641  A924642  A924643  A924644  A924645  A924646  A924647  A924648  A924650  A927827  A927828  A927829  A927830  A927831  A927832  A927833  A927834  A927835  A927836  A927837  A927838  A927839  A927840  A927841  A927842  A927843  A927844  A927845  A927847  A927848  A929704  A929705  A929706  A929707  A929708  A929710  A929711  A929713  A929714  A929716  A929717  A929893  A929894  A929896  A929897  A929898  A929901  A929902  A929903  A929904  A929906  A929907  A932604  A932605  A932607  A932609  A932610  A932611  A932612  A932613  A933706  A933707  A933709  A933710  A933781  A933782  A933783  A933784  A933785  A933786  A933787  A933788  A933789  A934499  A934500  A934503  A934504  A934505  A934506  A935182  A935184  A937237  A937238  A937239  A937240  A937241  A937645  A937646  A937647  A937648  A937649  A937650  A937651  A937652  A937653  A937654  A938289  A938290  A938291  A938293  A939611  A941306  A941307  A941308  A941309  A941310  A942157  A942159  A942160  A942161  A942162  A942163  A942164  A942165  A942166  A944241  A944242  A944244  A944245  A944246  A944247  A944249  A944250  A944251  A944253  A944254  A944255  A944256  A944257  A945204  A945205  A945438  A945441  A945445  A945447  A945450  A945451  A945452  A945454  A945455  A946808  A946811  A946813  A946815  A946816  A946817  A947150  A947151  A947152  A947153  A947154  A947155  A947396  A947397  A947398  A947399  A947400  A947401  A947402  A947403  A947405  A947407  A947408  A948123  A948124  A948125  A948126  A948127  A948128  A948129  A948130  A948131  A949486  A949487  A949488  A949489  A949490  A949491  A949492  A949493  A949494  A949495  A949496  A949497  A950324  A950329  A950332  A950333  A950335  A950337  A950341  A950344  A950346  A950753  A950754  A950945  A950946  A950947  A952499  A952500  A953289  A953290  A953291  A953292  A953294  A953295  A953296  A953297  A955938  A955939  A955940  A955943  A955944  A955945  A955946  A955947  A955948  A956502  A956504  A956505  A956506  A956507  A956508  A956509  A956510  A956511  A956512  A957173  A957175  A957176  A957270  A957271  A957272  A957273  A957274  A957282  A957461  A957470  A957472  A957473  A957475  A957476  A957477  A957478  A957479  A957480  A957481  A957483  A957484  A957485  A957486  A957488  A957490  A957492  A958533  A958534  A958535  A958536  A958537  A958538  A958539  A958540  A958542  A958543  A958547  A959339  A959341",
      "more_code_info": ""
    }
  ]
}