  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Belgrade",
      "address_1": "664 Cruiser Ln",
      "reason_for_recall": "The firm received one customer complaint concerning three OsteoSelect Demineralized Bone Matrix (DBM)  devices which contained 5.0cc of putty, instead of 10.00cc of putty as labeled.",
      "address_2": "",
      "product_quantity": "26",
      "code_info": "Lot/Serial numbers: B180228-751 through B180228-779;Model/Product Code: 359100;the expiration date of these devices is 2021-06-05.",
      "center_classification_date": "20200319",
      "distribution_pattern": "US  Nationwide Distribution in the states of  AZ,CA, NC, NJ, NY, PA, SC, TX, and WA. There were no OUS consignees.",
      "state": "MT",
      "product_description": "OsteoSelect Demineralized Bone Matrix (DBM) Putty 10.0cc in a syringe; STERILE R, Product Code: 359100, Dimensions: ONE(1) PACKAGE - CONTAINING 10.0cc. UDI: 00849777003681 - Product Usage: OsteoSelect DBM Putty is indicated for use as a bone void filler and bone graft substitute for voids or gaps that are not intrinsic to the stability of the bony structure. OsteoSelect DBM Putty is indicated for treatment of surgically created osseous defects or osseous defects from traumatic injury to the bone. OsteoSelect DBM Putty can be used as follows: - Extremities - Posterolateral spine - Pelvis.",
      "report_date": "20200325",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "XTANT MEDICAL INC",
      "recall_number": "Z-1526-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "85036",
      "termination_date": "20201217",
      "more_code_info": "",
      "recall_initiation_date": "20190503",
      "postal_code": "59714-9719",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}