  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Houston",
      "address_1": "100 Cyberonics Blvd",
      "reason_for_recall": "The pulse generators have a lower battery longevity than specified in their design requirement as a result of the devices being inadvertently left in a programmed ON state during manufacture.",
      "address_2": "",
      "product_quantity": "13 units",
      "code_info": "Serial Numbers: 37627, 38473, 38539, 41446, 42887, 43491, 44186, 44339, 44391, 44443, 44455, 44510, 50715",
      "center_classification_date": "20150427",
      "distribution_pattern": "AR, TX, GA, CA, MI, IN, FL, AZ, NV",
      "state": "TX",
      "product_description": "VNS Therapy DemiPulse Generator Model 103 and VNS Therapy DemiPulse Duo Generator Model 104; CYBERONICS, INC.",
      "report_date": "20150506",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Cyberonics, Inc",
      "recall_number": "Z-1526-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70804",
      "termination_date": "20151009",
      "more_code_info": "",
      "recall_initiation_date": "20150306",
      "postal_code": "77058-2069",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}