  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "West Sacramento",
      "address_1": "2040 Enterprise Blvd",
      "reason_for_recall": "Siemens investigation confirmed that a portion one lot of  panel Pos Combo 33 lot was packaged without desiccants.  The substrates on a panel packaged without a desiccant may deteriorate over time due to exposure to additional moisture in the panel pouch.",
      "address_2": "",
      "product_quantity": "664 panel boxes",
      "code_info": "Catalog number B10170211; Siemens Material number (SMN): 1071530, lot number 2014-09-12.: Expiration date 2014-09-12.",
      "center_classification_date": "20150427",
      "distribution_pattern": "Nationwide Distribution including Puerto Rico and to the states of : CT, GA, NC, OK, NY, SC, KY, TN, CA, WA, OR, NJ, OH, PA, FL, MI, TX, WI,  MO and NH.",
      "state": "CA",
      "product_description": "MicroScan Pos Combo Panel Type 33 (PC33):  Catalog number B1017-211     MicroScan Pos Combo 33 B1017-211 panels are designed for use in determining quantitative and /or  qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing  aerobic and facultative anaerobic gram-positive bacteria.",
      "report_date": "20150506",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Beckman Coulter, Inc.",
      "recall_number": "Z-1524-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70904",
      "termination_date": "20150427",
      "more_code_info": "",
      "recall_initiation_date": "20140408",
      "postal_code": "95691-3427",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}