      "status": "Terminated",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67815",
      "recalling_firm": "Maquet Datascope Corp - Cardiac Assist Division",
      "address_1": "1300 MacArthur Blvd.",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and approximately 104 countries.",
      "recall_number": "Z-1523-2014",
      "product_description": "Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump.     The Intra-Aortic Balloon Pump (IABP) is an electromechanical system used to inflate and deflate intra-aortic balloons in cardiovascular procedures.",
      "product_quantity": "11,881 units (US 5183 and OUS 6698)",
      "reason_for_recall": "Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.",
      "recall_initiation_date": "20140321",
      "center_classification_date": "20140521",
      "termination_date": "20180320",
      "report_date": "20140528",
      "code_info": "IABP Model Numbers  System 98 0998-00-0446-xx 0998-UC-0446-xx  System 98XT 0998-00-0479-xx 0998-UC-0479-xx  CS100i 0998-UC-0446Hxx 0998-UC-0479Hxx  CS100  0998-00-3013-xx 0998-UC-3013-xx  CS300 0998-00-3023-xx 0998-UC-3023-xx"
    }
  ]
}