  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73593",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Nationwide Distribution-including DC, and Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT,  DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.",
      "recall_number": "Z-1519-2016",
      "product_description": "Siemens SOMATOM Definition AS; Intended to produce cross-sectional images of the body.  Computed tomography (CT) x-ray system.",
      "product_quantity": "1096",
      "reason_for_recall": "Software bugs in these SOMATOM systems could possibly cause scan aborts, rescans of patients with additional dose, delayed diagnosis and as worst case scenarios, could possibly cause the need of additional contrast medium, accidental x-radiation to the user or impede customer workflow.",
      "recall_initiation_date": "20160302",
      "center_classification_date": "20160419",
      "termination_date": "20161213",
      "report_date": "20160427",
      "code_info": "Model Number: 8098027 SOMATOM Definition family of CT systems running on software versions VA44A_SP3a, VA44A_SP3b, VA44A_SP3c."
    }
  ]
}