      "country": "United States",
      "city": "Tarrytown",
      "address_1": "511 Benedict Ave",
      "reason_for_recall": "Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.",
      "address_2": "",
      "product_quantity": "18,489 units",
      "code_info": "Lots 260 through 275",
      "center_classification_date": "20140425",
      "distribution_pattern": "Worldwide distribution.",
      "state": "NY",
      "product_description": "IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diagnostics Products Ltd.   For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum.",
      "report_date": "20140507",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics",
      "recall_number": "Z-1518-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67210",
      "termination_date": "20180518",
      "more_code_info": "",
      "recall_initiation_date": "20131220",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}