  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Hartland",
      "address_1": "900 Walnut Ridge Dr",
      "reason_for_recall": "Users can merge a device import file with an image study that already has a confirmed report, which may result in including information not present at the time of physician interpretation.",
      "address_2": "",
      "product_quantity": "89 sites potentially have the affected software",
      "code_info": "Versions 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, and 9.0.8",
      "center_classification_date": "20170321",
      "distribution_pattern": "US Nationwide Distribution.",
      "state": "WI",
      "product_description": "Merge Cardio software.      Product Usage:  Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.",
      "report_date": "20170329",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Merge Healthcare, Inc.",
      "recall_number": "Z-1517-2017",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "76719",
      "termination_date": "20190808",
      "more_code_info": "",
      "recall_initiation_date": "20150922",
      "postal_code": "53029-8347",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}