  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Tarrytown",
      "address_1": "511 Benedict Ave",
      "reason_for_recall": "Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.",
      "address_2": "",
      "product_quantity": "5953 units",
      "code_info": "Lots 326 through 333",
      "center_classification_date": "20140425",
      "distribution_pattern": "Worldwide distribution.",
      "state": "NY",
      "product_description": "IMMULITE /IMMULITE 1000 Systems TIE Total IgE, REF LKIE1, SMN 10380867, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum.",
      "report_date": "20140507",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics",
      "recall_number": "Z-1517-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67210",
      "termination_date": "20180518",
      "more_code_info": "",
      "recall_initiation_date": "20131220",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}