  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Lexington",
      "address_1": "91 Hartwell Ave",
      "reason_for_recall": "TruFreeze Console caused a higher rate of liquid nitrogen (cryogen) to be delivered  and may cause: stricture,scarring, bradycardia, or pneumothorax",
      "address_2": "",
      "product_quantity": "82 units",
      "code_info": "Serial Numbers:   01-00106 through 01-00201",
      "center_classification_date": "20150423",
      "distribution_pattern": "Nationwide",
      "state": "MA",
      "product_description": "CSA Medical truFreeze Console- Cryosurgical Unit Cryogenic Surgical Device  Model: CC3-01",
      "report_date": "20150429",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "CSA Medical",
      "recall_number": "Z-1513-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "70851",
      "termination_date": "20150821",
      "more_code_info": "",
      "recall_initiation_date": "20150325",
      "postal_code": "02421-3137",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}